0% found this document useful (0 votes)
226 views15 pages

Manual

The document provides instructions for operating an analytical balance, including how to: 1) Fix the balance pan and remove the transport protection device. 2) Calibrate the balance using single point or linear calibration methods. 3) Operate the balance in weighing, counting, and percentage modes. 4) Set parameters like the unit of measure, automatic shutoff, and baud rate.

Uploaded by

diego
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
226 views15 pages

Manual

The document provides instructions for operating an analytical balance, including how to: 1) Fix the balance pan and remove the transport protection device. 2) Calibrate the balance using single point or linear calibration methods. 3) Operate the balance in weighing, counting, and percentage modes. 4) Set parameters like the unit of measure, automatic shutoff, and baud rate.

Uploaded by

diego
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 15

Contents

Balance-pan fixing

---------------------------------------------------

Protection device

---------------------------------------------------

Installation mode

---------------------------------------------------

Description of keys

---------------------------------------------------

Operation of weight
single point calibration

---------------------------------------------------

Operation of weight linear --------------------------------------------------calibration

Setting method

---------------------------------------------------

RS-232 serial interface

---------------------------------------------------

11

Operation instruction of
USB interface

---------------------------------------------------

13

Unit conversion table

---------------------------------------------------

14

JADEVER SCALE

Procedure of balance-pan fixing

In installing the balance pan,


the fixing screw in the center
of the plastic balance pan
should be tightened.

Then loose the transport


protection rod on the side of
the balance body.

Procedure of disassembling the protection device

1. There is a transport protection rod device on the left side of the


balance when facing the front of SNUG III analytical balance.

2. Before using SNUG III analytical balance, turn the protection rod
SNUG III

http://www.jadever.com.tw

JADEVER SCALE

counterclockwise to loose it and then release.

3. Before transporting SNUG III analytical balance, press the


protection rod inward and turn it clockwise to tighten.

Installation mode

1. When conducting the weight calibration of the product, pay attention to the
environment and avoid wind and vibration.

2. Usage of unsuitable cells or wrong connection of the wire may bring danger.

Description of keys

ON/OFF

ZERO

SNUG III

Key to power on and power off the balance.

Key for zeroing, which will function only within 2% of the maximum
weighing capacity; when in TARE mode, if the tare weight is smaller
than 2% of the maximum weighing capacity, the tare can be
cancelled and it can be zeroed; if the tare weight is greater than 2%
of the maximum weighing capacity, the tare can be cancelled.

http://www.jadever.com.tw

JADEVER SCALE

TARE

Key for deducting the weight; to deduct the weight of articles on the
balance.

MODE Key for mode shifting. The mode shifts in the sequence of weighing,
counting, percentage and printing time.

UNIT

SMPL

Weighing unit shifting key (13 kinds of units)

In the weighing mode, it is the key for Backlight switch (valid only
when Light is set to ON or OFF); in counting and percentage modes,
it is the key for sampling, press it continuously, you can select from
five sampling numbers- 20, 50, 100, 200 and 500.(Capable of
automatic averaging ).

Weighing
Press ON/OFFkey the screen will fully display and count down. One minute
later, the screen will display
0.00 X , and the weighing can begin. If backlight
is needed, please press SMPLkey.
Counting
Press MODEafter the balance is powered on, the screen will display
XXXX pcs, , then pressSMLPkey, the screen will display SE 0pcs .
Continuously pressTAREkey to select from the five sampling numbers -20, 50,
100, 200 and 500. After the selection, put the articles to be sampled on the
balance pan, pressSMLPkey and you can begin counting when a beep is
heard.
Percentage
Press MODEkey after the balance is powered on, the screen will display
XXXX % , ,then pressSMLPkey, the screen will display SE 0 % .
Continuously pressTAREkey to select from the five sampling numbers -20, 50,
100, 200 and 500. After the selection, put the articles to be sampled on the
balance pan, pressSMLPkey and you can begin calculating percentage when
SNUG III

http://www.jadever.com.tw

JADEVER SCALE

a beep is heard.
Print time setting:
Press MODEkey after the balance is powered on, the screen will display time
XXXX T ,then continuously press UNITkey to shift to month XX.XX D and
year XXXX Y. Press TAREkey to select the setting position , press SMPL
key to modify and press MODEkey to exit.
Notes: 1. The minimum sample weight 9 resolution. (For example, the sample
weight for 600g shall not be smaller than 0.09g.)
2. When the weight is unstable, the unit flashes.
3. The balance will warm up for five minutes after it is powered on.
4. When the weight is greater than full load + 9e, Err 5 will display.

Operation of weight single


point calibration

Step1: Please first power off the machine, press and hold down MODEkey ,
then press ON/OFF to power on the machine, release MODE key and
the screen will display as follows:

Count

Step 2: Press MODEkey once again and the screen will display as follows:

CAL

Step 3: Press UNITkey to begin the zero calibration, please do not put any
article on the pan, the screen will display as follows:

SNUG III

http://www.jadever.com.tw

JADEVER SCALE

ZEro

Please wait for a while until the


screen displays

on1

Step 4: PressTAREkey to select the calibration point, on1 is 1/3 load and on2
is 2/3 load and on3 is full load. Put the corresponding weight on the pan after
selection, wait for the beep, and the screen will display as follows:
PASS

Step 5: The single point calibration is now completed, clear the pan. Press
MODEkey and then pressSMPLkey to enter countdown mode.

Operation of weight linear


calibration

Step1: Please first power off the machine, press and hold down ZEROand
TAREkeys, then press ON/OFF to power on the machine, release
ZEROand TARE keys and the screen will display as follows:

SNUG III

http://www.jadever.com.tw

JADEVER SCALE

L-CAL

Step 2: Clear the pan, then press TAREkey and the screen will display as
follows: (zero calibration)

on0

Step 3: After a beep, the screen will display as follows:

on1

Step 4: Put the 1/3 load on the pan, after a beep , the screen will display as
follows:

on2

Step 5: Put the 2/3 load on the pan, after a beep, the screen will display as
follows:

on3

Step 6: Put the full load on the pan, after a beep , the screen will display as
follows:

PASS

SNUG III

http://www.jadever.com.tw

JADEVER SCALE

Step 7: The linear calibration is completed. Clear the pan and then pressTARE
key to enter countdown mode.

The corresponding loading values in the linear calibration for various


models of SNUG III balances
MODEL
Linear
calibration

SNUG
III-150

SNUG
III-300

SNUG
III-600

SNUG
III-1500

SNUG
III-3000

50g

100g

200g

500g

1000g

100g

200g

400g

1000g

2000g

150g

300g

600g

1500g

3000g

on1
Linear
calibration
on2
Linear
calibration
on3
Calibration Table

Setting method
Press and hold down theMODEkey , then pressON/OFFkey, release
MODEkey to enter the setting status:
MODEkey: for function shifting.
UNIT key: selecting key for enter and exit a function and function setting;
SMPLkey: for selection of function setting and for ending the setting.

SNUG III

http://www.jadever.com.tw

JADEVER SCALE

LCD display

Function
Description

MODEkey

Count

Function setting
selection

Function
setting
selection

UNIT key

SMPL

internal value
displays

View content

Note 1 on
Page 10

CAL

Single point
automatic
calibration

Page 4

unit

gctlbozdr
Selection of unit GN ozt dwt
MM tl.j tl.T
tl.Ht

ON/OFF

Note 2 on
Page10

InitU

gctlbozdr
Selection of unit
GN ozt dwt
when powering on
MM tl.j tl.T
tl.Ht

Note 3 on
Page 10

SNUG III

http://www.jadever.com.tw

JADEVER SCALE

Auto

Selection of
automatic power
Au-no/5/15/30/60
off

Note 4 on
Page 10

bAud

Selection of
transmission
speed

ZEro

Selection of
automatic zero
range

2400/4800/9600

Note 5 on
Page 10

d0/1/2/3/4/5

Note 6 on
Page10

Fil1/2/4/8

Note 7 on
Page11

Fil

Selection of
vibration
prevention

LigH

Selection of
backlight function

onoFF/oFF/ON

Selection of
printing mode

St/Co

Note 8 on
Page11

Print

Note 9 on
Page11

SNUG III

http://www.jadever.com.tw

JADEVER SCALE

Pr Fu

External
equipment

PC/ SH-24

Note 10 on
Page11

1. When the Count is displayed on the screen, pressUNITkey, you can


see the present internal value. Press UNITkey again and then pressMODE
key to return to the display of Count .
2. When
unit
is displayed on the screen, pressUNITkey to enter the
mode of selection of unit . The screen will display on g or
oFF g whereas
on represents the use of this unit and oFF is represents the disuse of this
unit. Press SMPLkey to shift between ON and OFF set on for the units to
be used and oFF for the units not to be used. Press UNITkey, you can
select from the thirteen units. When the selection is completed , press MODE
again to return to the display of unit
.
3. When InitU is displayed on the screen, pressUNITkey to enter the
mode of selection of unit when powering on. InitU xx X represents the
selection of this unit for use when powering on. PressUNIT key to shift among
the units until the desired unit is selected. Then, pressMODEkey to return to
the display of
InitU
.
4. When
Auto
is displayed on the screen, pressUNITkey to enter the
mode of selection of automatic power off. The screen displays XXX
X
represents the selection of automatic power off. PressUNITkey to shift until
the desired automatic power off is selected. Then, pressMODEkey to return to
the display of
Auto . Au-no-means not to select automatic power off, while
5-means 5 minutes, 15- 15 minutes, 30- 30minutes and 60- 60 minutes
(Automatic power off is not executive at unstable state).
5. When
bAud is displayed on the screen, pressUNITkey to enter the
mode of selection of setting the RS-232 transmission speed. The screen
displays XXXX
.X represents the RS-232 transmission speed. Press UNIT
to shift until the desired RS-232 transmission speed is selected. PressMODE
key again to return to the display of bAud
.2400, 4800, 9600
6.When ZEro
is displayed on the screen, pressUNITkey to enter the
mode of selection of automatic zeroing range. The screen displays dX X
represents the automatic zeroing range. Press UNITkey to shift until the
SNUG III

10

http://www.jadever.com.tw

JADEVER SCALE

desired automatic zeroing range is selected. PressMODEkey again to return


to display of
ZEro
d0, 1, 2, 3 , 4, 5( the higher the value, the larger the
automatic zeroing range is ).
7. When
FiL
is displayed on the screen, pressUNITkey to enter the
mode of selection of the vibration prevention. The display of FiL X
X
represents the vibration prevention grade. Press UNITkey to shift until the
desired vibration prevention grade is selected. PressMODEkey again to return
to the display of
FiL
FiL1, 2, 4, 8 ( the higher the value, the higher the
vibration prevention grade is ).
8. When
LigH is displayed on the screen, pressUNITkey to enter the
mode of selection of backlight function setting. The screen displays XXXX
X
represents the backlight function setting. Press UNIT key to shift until the
desired backlight function is selected. PressMODEkey again to return to the
display of
LigH
onoFF-is automatic backlight, oFF-is without backlight
and ON- is backlight.
9. When Print
is displayed on the screen, pressUNITkey to enter the
mode of selection of printing method. The screen displays
Pr-XX
X
represents the printing method. Press UNITkey to shift until the desired
printing method is selected. PressMODEkey again to return to the display of
Print
. Pr-Co-is continuous printing and Pr-St stable printing
10. When Pr Fu
is displayed on the screen, pressUNITkey to enter the
mode of selection of external equipment. The screen displays XXXX
X
represents the external equipment. Press UNITkey to shift until the desired
external equipment is selected. PressMODEkey again to return to the display
of
Pr Fu . -PCis to connect to the computer and SH-24 is to connect to
the dotted micro-printer
11. If it is not to enter the mode of function setting, pressSMPLkey to end the
function setting and return to normal weighing mode. Press MODEkey
continuously to shift between the function setting options in rotation.

RS-232 serial interface


The 9PIN connector at the right back of SNUG III balance is a RS-232 standard
SNUG III

11

http://www.jadever.com.tw

JADEVER SCALE

interface, with No. 2PIN for output, No. 5PIN for ground and others useless.

BAUD RATE2400 OR 4800 OR 9600 bps


DATA BIT 8
PARITY BITNNONE
STOP BIT1
CODEASCII

BIT FORMAT
LSB
0

MSB
1

START BIT

DATA BIT

STOP BIT

DATA FORMAT
N/G/T . W .

+/SP

LF CR
WEIGHT

SP

UNIT

N.Wnet weight
G.Wgross weight
T.Wtare weight
SPspace
Example:
2007-04-10 08:32:46
SNUG III

12

http://www.jadever.com.tw

JADEVER SCALE

T. W.: +

0.00g

N.W.: +499.98g
G.W.: +499.98g
Computer program10 OPENCOM1:9600,N,8,1,RS,DS,LFAS #1
(BASICA) 20 INPUT #1,A$
30 PRINT A$
40 GOTO 20
50 END

Operation instruction of USB interface

Precautions for using USB device:


1. This USB interface device can only communicate with computer.
2. The computer should be installed with the driver of the USB interface
equipment which can be downloaded from the companys website.
3. After the driver is installed, the communication between this device and the
computer behaves as a serial communication, and the serial transmission rate
should be set correspondingly.
The operation of USB interface equipment is as follows:
1. Use USB line to connect the electronic balance and the computer. Turn on the
electronic balance. If USB driver is not installed on the computer, prompt
information will be displayed on the computer to prompt you that a new
hardware is found and its driver is needed to be installed.
2. After the driver is installed, use the hyper terminal of the computer to test if
there is data transmission. Procedures for opening the computer hyper terminal
is: start
all programs
accessory
communication
hyper
terminal. Enter the name, click confirm, select COM3 or COM4 and click
confirm. Select the corresponding serial transmission rate (If the balance is set
SNUG III

13

http://www.jadever.com.tw

JADEVER SCALE

at 9600, select 9600), click confirm, then the communication can be made.
3. The communication mode of USB is the same as that of RS-232. If user has
his own computer receiving terminal, it can be used for the communication.

USB interface is an optional device. User can select USB interface device
or RS-232 interface device according to his need.

Unit Conversion Table


1

ct

[MET.CARAT]

0.1999694 g

lb

[AVORIRDUPOIS POUND]

453.59237 g

oz

[AVORIRDUPOIS OUNCE]

= 28.349523125 g

dr

[AVOIRDUPOIS DRAM]

1.7718451 g

GN

[GRAIN](U.K)

0.06479891 g

ozt

[TROY OUNCE]

31.1034768 g

dwt

[PENNY WEIGHT] (U.K)

1.55517384 g

MM

[MOMME] (JPN)

3.749996 g

tl.j

[HONG KONG JEWELRY TAEL]

37.4290018 g

tl.T

[TAEL](TWN) (Vietnam)

37.49995 g

tl.H

[HONG KONG TAEL]

37.799375 g

[TOLA] (INDIA)

11.6638038 g

SNUG III

14

http://www.jadever.com.tw