Calibration Manual
Calibration Manual
CALIBRATION MANUAL
1. Purpose:
1.1. To establish and maintain a calibration system to control the accuracy of equipment that is used
for measuring student performance evaluation, particularly state or national credentials.
Including measuring equipment used for public health and safety purposes.
1.2. To prevent inaccuracy, detect deficiencies and provide for timely positive corrective action.
1.3. The instructor is responsible for overall control of the calibration system and designating
master tools and equipment that are subject to calibration. Instructional tools and equipment
may be checked against the master tools for accuracy, as the instructor deems necessary.
2. Procedure:
2.1. Master tools will be calibrated by means of certified standards or equipment that has been
calibrated using certified standards.
2.2. Master gauge blocks will be used to calibrate the optical comparator in the Tool and Die lab.
2.3. A list of controlled, calibrated master tools and equipment is maintained by the instructor.
Master tools are used to check instructional tools as the instructor deems is necessary.
2.4. Any gauge or tool is considered out of tolerance if it exceeds one reference mark or graduation,
or as evaluated by the instructor. The instructor shall be informed of any out-of-tolerance
condition found during use so that proper actions may be taken.
3. Calibration Frequency:
3.1. The Optical Comparator will be calibrated with the Master gauge blocks each year prior to the
start of testing.
3.2. Tool and Die Lab measuring tools will be calibrated in accordance with National Institute of
Metalworking Skills (NIMS) standards. Tools shall be calibrated at the beginning of any job
requiring NIMS standards, or when the instructor determines it should be calibrated due to
accuracy requirements for the specific job.
3.3. Culinary Arts food preparation equipment shall be calibrated for temperature accuracy every
other year. Records of calibration shall be maintained in the Facility Managers office.
3.4. National Occupational Competetency Testing Institute (NOCTI) tools are to be calibrated prior
to the test.
Created 4/18/2002
Revised 5/6/05
5. Responsibility:
5.1. The instructors shall be responsible to identify and appropriately mark master measuring
equipment used to check the accuracy of instructional measuring equipment.
6. Records:
6.1. Facility Manager shall log the calibration of Optical Comparator on the Equipment Calibration
Log. The frequency of calibration and dates indicate the calibration was performed. The
Facility Manager will maintain logs for 3 years. This record is maintained in the Facility
Managers office and is available for review.
6.2. Instructors shall list the ID and description of equipment on the Equipment Calibration Logs.
The frequency of calibration and dates indicate the calibration was performed. The Facility
Manager will maintain logs for 3 years. This record is maintained in the Facility Managers
office and is available for review.
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
Created 4/18/2002
It is extremely important that all parts of the micrometer be kept clean during
calibration.
Wipe micrometer clean with a clean cloth and apply a drop of fine oil to
spindle.
First you will need to remove play/looseness from the threads due to wear in
the spindle.
To do this back off the thimble.
Insert the C-spanner into the slot or hole of the adjusting nut.
Turn the adjusting nut clockwise until play between the threads has been
eliminated.
Do not over tighten, the spindle should move freely when the ratchet stop or
friction thimble is being turned.
Next you will need to test the accuracy of the micrometer.
You will need a gage block from the Master Gage Blocks for the size
micrometer that you will be checking.
Open the micrometer so that the gage block will fit between the measuring
faces.
Check to make sure that the measuring faces are clean and there are no chips
in them.
Place the master gage block between the measuring faces.
Turn the thimble using the ratchet stop or friction thimble until the measuring
faces contact the master gage block.
The reading of the micrometer must be the same +/- .0005 as the master gage
block.
If the reading is more than +/- .0005 you will need to adjust the micrometer.
To adjust the micrometer close the measuring faces carefully until they
contact the master gage block.
Insert the C-spanner into the hole or slot provided in the sleeve.
Carefully turn the sleeve until the index line on the sleeve coincides with the
zero line on the thimble.
Recheck the accuracy of the micrometer by opening the micrometer and then
closing the measuring faces carefully until they contact the master gage block
again.
The micrometer will be considered out of tolerance if after 3 tries the
micrometer will not hold the reading of the master gage block +/- .0005.
The micrometer will be considered out of tolerance if after repeating steps 10
thru 14 and the micrometer will not hold +/- .0005 it will then be removed
from service.
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
To check calibration of line gauge you will need to print out from Pagemaker 7.0
measurement gradations of 6, 12, 18, 24, 30, and 36 picas on a piece of paper.
2. Compare line gauge to measurements on the paper.
3. If your reading is more than 2 points at 30 picas you will need to discard the line
gauge because there is no way to recalibrate.
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Calibration * The prime standard of calibration is a factory calibrated Fluke model 111
These oscilloscopes will be calibrated once every two-school years.
.
For proper calibration please consult the attachment instructions, or refer to
TDS 200-series digital oscilloscope programmer manual on page 2-16.
Calibration checks will be performed annually or as necessary.
Created 4/18/2002
Revised 5/6/05
Calibration * The prime standard of calibration is a factory calibrated Fluke model 111
This meter will be a master tool and is maintained by the instructor.
1. For proper calibration please consult the attachment instructions (Specifically
Pg.13-14) Calibration checks will be performed annually.
2. The functions to be checked for calibration are: AC & DC volts, AC & DC
amps, resistance, capacitance, diode check and frequency.
3. To check calibration, connect the meter leads to a known source in tandem
with the factory calibrated Fluke 111 DMM. All functions of the meter being checked
must be compared to the factory calibrated multimeter and be within the + or - 3%
tolerance range.
4. If the meter is found to be in tolerance for all checked functions, an updated
calibration sticker will be applied.
5. The digital multimeter will be considered out of tolerance if after 3 tries the
digital multimeter will not hold a + or 3% value when compared to the factory calibrated
meter. If the reading of the meter being checked is more than + or 3% of the reading on the
factory calibrated meter, the meter being checked will be taken out of service and sent to the
Fluke Corp. for factory calibration.
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
Created 4/18/2002
Revised 5/6/05
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
To calibrate the gage start the display so it reads in inches and showing SET mode,
slide the moving arm to the left so that it engages the positive Stop on the fixed arm.
Dress the zero (button).
Open the arms and set the pointed anvils onto the gauger service of the calibration
gage.
The display should now read 7.000 +/- .0005.
If it does not you will need to adjust the mechanical calibration.
Place the pointed anvils onto the gauger surface of the calibration gage
Depress the zero (button).
Slide the moving arm to the left until it engages the Positive Stop on the fixed arm.
The reading on the display will now be a negative value, which is approximately 7.
Using the hex wrench provided, loosen the setscrew on the Positive Stop.
Turn the Positive stop in or out as required until the display reads precisely 7.000
(+/- .0005).
Press the zero (button) to zero the display.
Open the arms inset the anvils back into the calibration gage.
The reading should now be 7.000 +/- .0005.
When reading is correct set lock screw in the calibration set screw to lock the
calibration setting.
The digital brake drum gage will be considered out of tolerance if after 3 tries it does not repeat
the reading within +- .0005 remove from service.
Created 4/18/2002
Revised 5/6/05