Electrical Measurements Lab Manual
Electrical Measurements Lab Manual
com
www.jwjobs.net
Topic
Week
W1
(ii)
W2
W3
W4
W5
Unit-III(Measurement of Power)
4
www.jntuworld.com
W6
www.jntuworld.com
www.jwjobs.net
W7
W8
W10
W11
W9
11
12
W12
W14
W16
Experiment no.1
www.jntuworld.com
W13
W15
www.jntuworld.com
www.jwjobs.net
:
NAME
1-Phase Energy meter
Wattmeter(Sub-standard or calibrated)
Voltmeter
Ammeter
Rheostat
Single Phase Variac
Connecting wires
Stop watch
TYPE
AC
Dynamometer type, LPF
MI
MI
Digital
RANGE
230 V/10 A
10A, 300V
300 V
10A
50, 15A
0-270V, 10A
QTY.
1 no.
1 no.
1 no.
1 no.
1no.
1 no
1 set
1no.
CIRCUIT DIAGRAM:
THEORY :
Induction type of energy meters are universally used for measurement of energy in domestic and
industrial a.c. circuits. Induction type of meters possesses lower friction and higher torque/weight ratio.
Also they are inexpensive and accurate, and retain their accuracy over a wide range of loads and
temperature conditions.
There are four main parts of the operating mechanism:
(i)
Driving system
(ii)
Moving system
(iii)
Braking system and
(iv)
Registering system.
Driving System: The
driving system of the meter consists of two electro-magnets. The core of these
electromagnets is made up of silicon steel laminations. The coil of one of the electromagnets is excited by
the load current. This coil is called the current coil. The coil of second electromagnet is connected across
the supply and, therefore, carries a current proportional to the supply voltage. This coil is called the
pressure coil. Consequently the two electromagnets are known as series and shunt magnets respectively.
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
Copper shading bands are provided on the central limb. The position of these banks is adjustable. The
function of these bands is to bring the flux produced by the shunt magnet exactly in quadrature with the
applied voltage.
Moving System: This consists of an aluminium disc mounted on a light alloy shaft. This disc is positioned
in the air gap between series and shunt magnets.
Braking System: A permanent magnet positioned near the edge of the aluminium disc forms the braking
system. The aluminium disc moves in the field of this magnet and thus provides a braking torque. The
position of the permanent magnet is adjustable, and therefore, braking torque can be adjusted by shifting
the permanent magent to different radial positions as explained earlier.
Registering (counting) Mechanism: The function of a registering or counting mechanism is to record
continuously a number which is proportional to the revolutions made by the moving system.
In all induction instruments we have two fluxes produced by currents flowing in the windings of the
instrument. These fluxes are alternating in nature and so they produce emfs in a metallic disc or a drum
provided for the purpose. These emfs in turn circulate eddy currents in the metallic disc or the drum. The
breaking torque is produced by the interaction of eddy current and the field of permanent magnet. This
torque is directly proportional to the product of flux of the magnet, magnitude of eddy current and effective
radius R from axis of disc. The moving system attains a steady speed when the driving torque equals
braking torque.
The term testing includes the checking of the actual registration of the meter as well as the adjustments
done to bring the errors of the meter with in prescribed limits. AC energy meters should be tested for the
following conditions:
1.
At 5% of marked current with unity pf.
2.
At 100% (or) 125% of marked current.
3.
At one intermediate load with unity pf.
4.
At marked current and 0.5 lagging pf.
PROCEDURE :
1.
2.
3.
4.
5.
6.
7.
8.
9
10.
TABULAR FORM:
S.
No.
Voltmeter
(Volts)
Ammeter
Reading Amps
Wattmeter
Reading W
Time for
20 rev
www.jntuworld.com
Theoretical
kWH (O.R)
Practical
wxt (A.R)
%
error
www.jntuworld.com
www.jwjobs.net
O.R=OBSERVED READING
A.R=ACTUAL READING
%error
MODEL CALCULATIONS:
Observed reading = 1/(energy meter constant (k)/no. of revolutions)
Where , no. of revolutions = 10
energy meter constant k=900 rev/kwh
Actual reading
= W *t
%error = [(O.R-A.R)/A.R] *100
%correction = - % error = [(A.R-O.R)/A.R] *100
MODEL GRAPH :.
Load current
PRECAUTIONS :
1.
2.
3.
4.
5
6
7.
8.
9
10
11
12
RESULT :
The given single phase energy meter is tested at different loads and calibration curve is plotted.
QUESTIONS:
1.
2.
3.
4.
5.
6.
7.
8.
9.
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
10. The operating torque is directly proportional to speed, state true or false.
EXPERIMENT NO.2
CALIBRATION OF DYNAMOMETER POWER FACTOR METER
AIM :
Testing and Calibration of the given Dynamometer Power Factor Meter by using a sub-standard wattmeter,
ammeter and voltmeter.
APPARATUS :
SL.
NO.
1.
p.f. meter
2.
Wattmeter
3.
4.
5.
6
7
Voltmeter
Ammeter
Rheostat
Connecting wires
Single Phase Variac
NAME
TYPE
RANGE
QUANTITY
Dynamometer
type, UPF
Dynamo meter
type LPF
MI
MI
10A, 300V
1 no.
10A, 600V
1 no.
300 V
10A
50 Ohm/5A
AC
270V, 10A
2 no.s
1 no.
1no.
1 set
1no.
CIRCUIT DIAGRAM:
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
THEORY:
Power factor meters indicate directly, by a single reading, the power factor of the circuit to which they are
connected.
Power factor meters like wattmeters have a current circuit and a pressure circuit. The current circuit
carries the current (or definite fraction of this current) in the circuit whose power factor is to be measured.
The pressure circuit is connected across the circuit whose power factor is to be measured and is usually
split up into two parallel paths one inductive and the other non-inductive. The deflection of the
instrument depends upon the phase difference between the main current and the currents in the two paths of
the pressure circuit, i.e. upon the phase angle or power factor of the circuit. The deflection is indicated by
a pointer. The deflection of the instrument is a measure of phase angle of the circuit. The scale of the
instrument can be calibrated in directly in terms of power factor.
The moving system of power factor meters is perfectly balanced at equilibrium by two opposing forces and
therefore there is no need for a controlling force. Hence when a power factor meter is disconnected from a
circuit the pointer remains at the position which it occupied at the instant of disconnection.
The instrument must be designed for, and calibrated at, the frequency of the supply on which its to be used.
In case the meter is used for any other frequency or if the supply contains harmonics it will give rise to
serious errors in the indication on account change in the value of reactance of choke coil.
PROCEDURE :
1.
Connections are made as per the circuit diagram.
2.
With variac in minimum or zero output position, the Rheostat in maximum resistance position,
close the DPST switch.
3.
Apply rated voltage 230V to pressure coils of the P.F meter and wattmeter with the variac. .
4
Close the load DPST and apply a suitable load . Note down the readings of the instruments
www.jntuworld.com
www.jntuworld.com
5
6.
7.
8.
9
www.jwjobs.net
Increase the load in steps and in each step note down the readings of the instruments. Tabulate the
readings.
Gradually reduce the load in steps and open the load DPST.
Vary the variac gradually to minimum or zero output position and open the supply DPST.
The PF measured from PF meter under test and actual PF from the reading of wattmeter, voltmeter
and ammeter calculated and %error, %correction are found.
Draw the graph between Load current (vs) % Error.
TABULAR FORM:
S.No.
Voltmeter
Reading
(Volts)
Ammeter
Reading , I
(Amps)
Wattmeter
Reading, W
(Watts)
P.F. meter
Reading(S2)
COS
=W/VI
(S1)
%Error
=S1-S2*100
S1
%error
MODELGRAPH:
Lead p.f.
Lag p.f.
PRECAUTIONS :
1.
2.
3.
4.
5.
6.
7.
8.
9
10
11.
12.
13
14
15
16
17
RESULT :
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
The dynamometer type power factor meter is tested at different power factors by using phase-shifting
transformer and calibration curve is plotted.
QUESTIONS:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
Wattmeter(Watts)
0
50
Powerfactor
meter
1
1
%error=S1-S2/S1 *100
S1= cos=W/VI
S1=50/230=0.217
Powerfactor meter reading(S2)=1
%Error =0.217-1/0.217 *100
%error = -360.8
ii)
S1=100/230=0.434
S2=Powerfactor reading=1
%Error =0.434-1/0.434*100
%error = -130.4
www.jntuworld.com
cos=W/VI Error=
S1-S2/S1
0
0
0.217
-360.8
www.jntuworld.com
www.jwjobs.net
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
EXPERIMENT NO.3
CALIBRATION OF PMMC AMMETER AND VOLTMETER USING CROMPTON D.C.
POTENTIOMETER
AIM : To calibrate PMMC Ammeter and Voltmeter using Crompton DC potentiometer.
APPARATUS REQUIRED :
Sl.
NO.
1.
NAME
TYPE
DC Crompton
Potentiometer
Standard cell
Volt ratio box
DC
5.
Regulated power
supply(RPS)
Ammeter
6
7
8
Voltmeter
Connecting wires
Sensitive galvanometer
9
10
11
12
Standard resistance
Rheostat
battery
DC SUPPLY
2.
3.
4.
RANGE
QTY.
1 no.
DC
1 no.
1 no.
DC
MC
0-2A
1no.
MC
0-300V
Spot reflecting MC
type
30-0-30
1 no
1 set
1 no
1 no
1 no
1 no
1no.
CIRCUIT DIAGRAM :
CALIBRATION OF VOLTMETER
CALIBRATION OF AMMETER :
www.jntuworld.com
1 no.
www.jntuworld.com
www.jwjobs.net
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
THEORY :
A Potentiometer is an instrument designed to measure an unknown voltage by comparing it with a
known voltage. The known voltage may be supplied by a standard cell or any other known voltage
reference source. Measurements using comparison methods are capable of a high degree of accuracy
because the result obtained does not depend upon on the actual deflection of a pointer, as is the case in
deflectional methods, but only upon the accuracy with which the voltage of the reference source is known.
Another advantage of the potentiometers is that since a potentiometer makes use of a balance or null
condition, no current flows and hence no power is consumed in the circuit containing the unknown emf
when the instrument is balanced. Thus the determination of voltage by a potentiometer is quite independent
of the source resistance.
It can also be used to determine current simply by measuring the voltage drop produced by the unknown
current passing through a known standard resistance. The potentiometer is extensively used for a
calibration of voltmeters and ammeters and has in fact become the standard for the calibration of these
instruments. For the above mentioned advantages the potentiometer has become very important in the field
of electrical measurements and calibration.
Modern laboratory type potentiometers used calibrated dial resistors and a small circular wire of one or
more turns, thereby reducing the size of the instrument. The resistance of slide wire is known accurately,
the voltage drop along the slide wire can be controlled by adjusting the value of working current. The
process of adjusting the working current so as to match the voltage drop across a portion of sliding wire
against a standard reference source is known as Standardisation.
PROCEDURE :
STANDARDIZATION:
1.
2.
3.
4.
PMMC VOLTMER
1.
Volmeter under test is connected across potential divider in such a way that p.d across voltmeter
can be varied
2.
V.R box is used in parallel to potentiometer to reduce voltage to the range of potentiometer
3.
Change the function knob to E1 position. Switch on RPS and adjust a suitable voltage on V.R box
4
Press the galvano key of the potentiometer and adjust the slide contact and slide wire until the spot
reflecting galvanometer gives null deflection.
5
Note down the readings of voltmeter, and potentiometer slide contact and slide wire readings.
6
Repeat the steps 4 and 5 for different voltages from RPS
7.
Reduce the voltage of RPS and RPS to zero. Switch off the supply.
8
Draw the graph between Load current (vs) % Error.
NOTE :
1.
RPS must be at 2 volts only.
2.
Smoothly vary the knobs on the potentiometer.
3.
Keep the spot reflecting galvanometer switch to AC mains position and rotating key to free
position during the conduction of experiment.
4.
At the end of the experiment remove the supply to the galvanometer and move the rotating key to
lock position and short the galvanometer output terminals.
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
CALCULATION :
Calibration of voltmeter:
VR Box ratio = Output voltage / Input voltage
Vact
= Voltmeter reading
Vtrue
= Reading obtained from potentiometer
= [Coarse voltage + Fine voltage] x Ratio of volt-ratio box
Vact Vtrue
100
V
true
% error =
PROCEDURE:
STANDARDIZATION:
1.
2.
3.
4.
PMMC AMMETER:
1
Ammeter to be calibrated is connected in series with variable resistor R & standard resistance S
2
The standard resistance should of such a magnitude that current passed through it doesnt exceed
range of potentiometer
3
V.R box is used in parallel to potentiometer to reduce voltage to the range of potentiometer
4.
Keep the rheostat at maximum position and change the function knob to E1 position. Switch on
RPS(2) and apply 30 volts.
5
Vary the rheostat gradually and adjust suitable current.
6.
Press the galvano key on the potentiometer and adjust the slide contact and slide wire until the
spot reflecting galvanometer gives the null deflection.
7.
Note down the readings of ammeter, voltmeter, and potentiometer slide contact and slide wire
readings.
8.
Repeat the steps 5 to7 for different values of current.
9
Vary the rheostat to maximum position, reduce the voltage of RPS and RPS to zero. Switch off the
supply.
10
Draw the graph between Load current (vs) % Error.
TABULAR FORM:
FOR VOLTMETER:
S.No
Voltmeter
reading(V) , Vtrue
E2 value(V)
Vact=(R1+R2/R2)
*E2
%error
Vact Vtrue
100
Vtrue
FOR AMMETER:
S.No
Ammeter
reading(A)
E2 value(V)
www.jntuworld.com
Iact=E/R
%error
www.jntuworld.com
www.jwjobs.net
I act I true
100
I true
Itrue
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
CALCULATION :
CALIBRATION OF AMMETER:
Iact
= Ammeter reading
7.
RESULT :
The calibration of PMMC Voltmeter and Ammeter is done using Crompton DC potentiometer. The
calibration curves for voltmeter and ammeter are drawn.
QUESTIONS:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
Vact = (1+2 / 2) * 5
=7.5V
Vtrue=0.25+51mv
=0.301V
% Error=(7.5-0.301/0.301 ) *100
=23.9%
Calibration of Ammeter:
%Error= (Iact-Itrue) / Iact *100
Iact =E/R
Itrue = Ammeter reading (A)
Iact = E/R=5/2
=2.5A
Itrue= 0.1 amps
% Error=(2.5-0.1/0.1 ) *100
=24%
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
EXPERIMENT NO.4
KELVINS DOUBLE BRIDGE
AIM :
To determine the low resistance by setting up Kelvins double bridge.
APPARATUS REQUIRED :
Sl. NO.
1
2.
3.
4.
5.
6
NAME
Kelvins double bridge kit
Galvanometer
Rheostat.
DMM
Patch cards/ Connecting wires
RPS
TYPE
RANGE
5A,50 OHM
DIGITAL
DC
2V
QTY.
1 no
1 no.
1 no.
1 no.
1set.
1 no
CIRCUIT DIAGRAM:
THEORY :
X is the unknown resistance to be measured.
S is the standard resistance.
The ratio of Q/q must be equal to M/m under balance conditions; there is no current through
galvanometer G. The first of ratio arms is Q and M. The second set of ratio arms of m is used to connect the
galvanometer to a point at the appropriate potential b/w point e and f to eliminate the effect of connecting
lead of resistance r between then the unknown resistance X and standard resistance S.
If
Q M
Q
= , thenX = S
p m
M
PROCEDURE :
1.
2.
3.
4.
5.
www.jntuworld.com
www.jntuworld.com
6.
www.jwjobs.net
Note down the readings of main dial slide wire and multiplier.
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
OBSERVATIONS :
S.No.
un Resistance or practical
1.
2.
3.
4.
SAMPLE CALCULATIONS :
Unknown resistance = multiplier *(main dial reading + slide wire reading)
PRECAUTIONS :
1.
2.
3.
RESULT :
The unknown resistances of earth wires, having different parameters, are measured by using Kelvins
double bridge.
QUESTIONS:
1.
2.
3.
4.
5.
6.
7.
8.
9.
Classify resistance?
Examples of high resistance?
What are the methods employed in measuring low resistances?
Which is the most accurate method to measure the low resistances? State the reason?
Kelvin double bridge is a modified version of ?
What is the main problem in measuring low resistances?
How do we measure high resistances?
Practical methods to measure earth resistance?
What are the quantities that are measured by ac bridges?
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
Tolerance = 2.5
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
EXPERIMENT NO. 5
CT TESTING BY SILSBEES METHOD
AIM:
To measure the % ratio error and phase angle error of given CT by Silsbees method.
APPARATUS:
SL.
NO.
1.
2.
3.
4.
5.
CT
CT
Voltmeter
Ammeter
Wattmeter
6
7
8
9
10
Connecting wires
Rheostat
Loading Burden Rheostat
Phase-shifting transformer
Single Phase supply
NAME
TYPE
Precision (Standard)
Commercial
MI
MI
Electrodynamometer
Type LPF
RANGE
5/5
5/5
300 V
10A
10A, 600V
QUANTITY
1 no.
1 no.
1 no.s
2 no.
2no.
1 set
AC
50 Ohms, 5A
50 Ohms, 5A
3 Ohms, 420V, 10A
230V, 10A
1 no
1 no
1no.
CIRCUIT DIAGRAM:
THEORY :
Methods of experiment testing instrument transformers i.e., finding their ratio and phase angle errors may
be broadly classified into two groups. Absolute methods and comparison methods each of these test
methods can be classified according to measurement technique employed as Deflecting methods and Nullmethods.
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
Silsbees method is a comparison method. There are two types of Silsbees methods deflectional
and here the ratio and phase angle of the test transformer are determined, in term of a standard transformers
having the same nominal ratio.
The two transformers are connected with their primaries in series. An adjustable burden is put in
the secondary circuit of the transformers under test. An ammeter is included in the secondary circuit of the
standard transformers so that the current may be set to the desired value. W1 is a wattmeter whose current
coil is connected to carry the secondary current of the standard transformers. The current coil of wattmeter
W2 carries a current which is the difference between the secondary currents of the standard and test
transformers. The voltage circuits of the wattmeters lies their pressure coils are supplied in parallel from a
phase shifting transformer at a constant voltage V.
PROCEDURE:
1.
2.
3.
4.
5.
6.
7.
TABULATION:
S.No.
Ammeter
Reading
I1(Amps)
Ammeter
Reading
I2(Amps)
W1p
(Watts)
Rs =I1/I2
% Ratio error (Rx) = Rs {(W2p / W1p) +1}
Phase angle error (x) = W2q/ (W1p W2p)
W2p
(Watts)
W1q
(Watts)
+ s
FORMULAE USED :
Nominal ratio of standard CT = 10/5 = 2
Phase angle error of standard CT = 5
Actual ratio of standard CT = 1.99
W01
V I ss
Rs R
% ratio error =
Rx
W02 180
V
I
ss
PRECAUTIONS :
1.
Take care, while varying the phase angle with phase shifting transformer.
www.jntuworld.com
W2q
(Watts)
www.jntuworld.com
www.jwjobs.net
2
3
4.
5
6
7
8
9
10
11
RESULT :
The current transformer is tested by using Silsbees method and, the ratio and phase angle errors are
calculated.
QUESTIONS:
1
2
3
4
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
EXPERIMENT NO.6
A. MEASUREMENT OF CAPACITANCE BY SCHERING BRIDGE
AIM : To find the capacitance of unknown capacitor using Schering bridge.
APPARATUS :
Sl. NO.
1
2.
3.
4.
5.
6
7
NAME
Schering bridge circuit
Head phones
Decade capacitance box
DMM
Patch cards
RPS
Galvanometer
TYPE
RANGE
digital
230v
QTY.
1 no
1 no.
1 no.
1 no.
1set
1 no
1 no
CIRCUIT DIAGRAM:
THEORY:
The Schering bridge is very widely used for the measurement of capacitance, dielectric loss and power
factor of capacitors. The advantage of using this bridge is that it can be employed in both low voltage and
high voltage measurements.
From the circuit :
C1 is the unknown capacitance and
R1 is a resistance representing its loss component
C2 is a standard air capacitor and it is loss free
R3 & R4 are non-inductive standard resistors.
R3 is fixed and R4 is variable
C4 is a variable capacitors.
The detector may be a head phone or vibration galvanometer
E is a low voltage a.c. source r 230v,50hz,supply
The bridge is balanced by adjusting
C4 and R4
At balance, we have V1 = V2 & V3 = V4
1
1
V1 = r1 +
,V 2 = 1
,V3 = I1 R3 & V4 I 2 Z 4
jwc1
jwc 2
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
1
R 4
R4
jwc 4 =
Where Z4 =
1
jwR 4C 4 + 1
R4 +
jwc 4
12 R4
Or V4 =
1 + jwR 4 C 4
1
1
. (1) and
11 r1 +
= I 2
jwc1
jwc 2
I 2 R4
I1R3 =
.. (2)
1 + jwR 4 C 4
From (1) & (2)
1
jwC1 1 + jwR4 C 4
r1 +
R3 = ( jwc 2 )( R4 ) or
R4
1
r1 +
jwg 1 + jwC 4 R4
1
jwc 2
= R3
R
jR4 R3 R4 C 4
j 3
wC1 = C 2
wC 2
Simplifying, r1R4 Equating real and imaginary parts
Separately, we get :
R3 R4 C 4
C2
r1R4=
& C1= Cx =(R4/R3) x C2
NOTE: C1 =Cx1= Cx2=
C2=Cs1=Cs2=.
PROCEDURE :
1.
Connections are made as per the circuit diagram.
2.
Connect the unknown capacitance Cx between the terminals A & D.
3.
Switch on the supply and vary the Resistance knob (R 1) until minimum sound is obtained from the
loud speaker.
4.
Switch off the supply and measure the Resistance (R1) between C & B by using DMM.
5.
Repeat the above steps for different values of capacitance & tabulate the readings.
6.
Switch off the supply.
PRECAUTIONS :
1.
Avoid loose connections.
2.
Resistance should be varied very smoothly.
3.
Switch off the supply when the resistance R1 is measured.
FORMULA USED:
R3 R 4 C 4
r1R4=
& C1= Cx =(R4/R3) x C2
C2
NOTE: the value of unknown capacitors is
C = 0.01F
0.02 F
0.03 F
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
0.04 f
s.no
Cs
(known
capacitance)
R4
www.jntuworld.com
R3
Cx=CsR4/R3
www.jntuworld.com
www.jwjobs.net
QUESTIONS:
1. What do you mean by high voltage Schering Bridge?
2. State some of the errors that occur in bridge measurements?
3. Anderson Bridge is a modified version of __________
4. In Anderson Bridge the self inductance is measured in comparison with ___________
5. What are the resistors need to be adjusted to get the balance
6. At what condition the galvanometer detector will be replaced by the head phone.
7. Schering bridge is used for the measurement of ______________
8. What is meant by loss angle ?
9. Why we are doing electrostatic shielding for high voltage Schering bridge?
10. What are the elements need to be adjusted to obtain balance in Schering bridge?
RESULT : the value of unknown capacitance is calculated by using Schering bridge
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
Calculation:
By taking C2 value as 0.1f
We get R4 = 10.3 k
R3 = 560m
Cx = (R4 C2)/ R3
Then, Cx = (10.3 k*0.1f)/ 560m
= 1.83pf
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
NAME
Andersons bridge circuit
Head phones
Decade inductance box
DMM
Patch cards
RPS
Galvanometer
TYPE
RANGE
DIGITAL
230
QTY.
1 no
1 no.
1 no.
1 no.
1set
1 no
1 no
CIRCUIT DIAGRAM:
THEORY :
Andersons bridge is a modification of the Maxwells inductance capacitance bridge. In this method, the
self-inductance is measured in terms of a standard capacitor. This method is applicable for precise
measurement of self-inductance over a very wide range of values.
Figure shows the connections and the phasor diagram of the bridge for balanced conditions:
Let
L1 = Self-inductance to be measure
R1 = resistance of self-inductor,
r1 = resistance connected in series with self-inductor,
r, R2, R3, R4 = known non-inductive resistances, and
C = fixed standard capacitor.
At balance, I1 = I3 and I2 = Ic + I4
1
Ic = I1jCR3.
Now
I1R3 = Lc x
j C
Writing the other balance equations
1
= (I2 Ic) R4.
Ic r +
jC
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
1
= (I2 IjCR3)R4 or I1(jCR3r + jCR3R4 +R3) = I2R4 (ii)
jCR3 I1 r +
jC
+
+ jCR3 R2
I1 (r1 + R1 + jl1 jCR3r) = I1
R4
R4
R2 R3
r1
R4
R3
[r(R4 + R2) + R2R4]
R4
An examination of balance equations reveals that to obtain easy convergence of balance, alternate
adjustments of r1 and should be done as they appear in only one of the two balance equations.
and L1 = C
ADVANTAGES:
1. In case adjustments are carried out by manipulating control over r1 and r, they become independent of
each other. This is a marked superiority over sliding balance conditions met with low Q coils when
measuring axwells bridge. A study of convergence conditions would reveal that it is much easier to
obtain balance in the case of Andersons bridge than in Maxwells bridge for low Q-coils.
2. A fixed capacitor can be used instead of a variable capacitor as in the case of Maxwells bridge.
3. This bridge may be used for accurate determination of capacitance in terms of inductance.
DISADVANTAGES:
1. The Andersons bridge is more complex than its prototype Maxwells bridge. The Andersons bridge
has more parts and is more complicated to set up and manipulate. The balance equations are not simple
and in fact are much more tedious.
2. An additional junction point increases the difficulty of shielding the bridge.
Considering the above complications of the Andersons bridge, in all the cases where a variable capacitor is
permissible the more simple Maxwells bridge is used instead of Andersons bridge.
PROCEDURE :
1.
Connections are made as per the circuit diagram with an audio oscillator and head phones
connected to proper terminals of the Andersons bridge.
2.
Connect the unknown inductor L as shown in the circuit diagram.
3.
Switch on the supply and select a certain value of C say 0.01 F.
4.
Adjust R1and r1alternately till the head phones give minimum or no sound.
5.
Note down the values of S, M and C at this balanced condition.
6.
Repeat steps (4) and (5) for the same inductance by selecting different value of C.
7.
Repeat the above steps for different values of unknown inductance.
8.
Switch off the supply.
NOTE :
1.
The value of C is so chosen that there is sufficient adjustment available in the value of M.
2.
When C is small, M will be large.
3.
The bridge is useful for measuring small values of inductor such as 50, 100, 150 and 200 mH.
Note the value of unknown inductances
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
1.
2.
10mH
100mH
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
CALCULATION :
L value is calculated by the given formula.
L1 = C
R3
[r1(R4+ R2) + R2R4]
R4
R2 R3
r1
R4
R1 =
S.NO
C(KNOWM
r1
R1
R2
R3
R4
L1
CAPACITANCE)
RESULT :
The value of unknown inductance is calculated by using Anderson Bridge
QUESTIONS:
1. What do you mean by high voltage Schering Bridge?
2. State some of the errors that occur in bridge measurements?
3. Anderson Bridge is a modified version of __________
4. In Anderson Bridge the self inductance is measured in comparison with ___________
5. What are the resistors need to be adjusted to get the balance
6. At what condition the galvanometer detector will be replaced by the head phone.
7. Schering bridge is used for the measurement of ______________
8. What is meant by loss angle ?
9. Why we are doing electrostatic shielding for high voltage Schering bridge?
10. What are the elements need to be adjusted to obtain balance in Schering bridge?
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
R2 R3
r1
R4
R3
Calculation:
In Andersons Bridge we have to calculate the unknown inductance value by using the
formula
R3
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
EXPERIMENT NO.7
MEASUREMENT OF 3-PHASE REACTIVE POWER WITH SINGLE PHASE WATTMETER
AIM : To measure the 3-Phase reactive power using single wattmeter.
APPARATUS
Sl. NO.
1.
2.
3.
4.
5
6
NAME
Wattmeter
Voltmeter
Ammeter
3 Phase variable inductive load
TYPE
Dynamometer type, LPF
MI
MI
RANGE
10A, 300V
300 V
10A
0-10A, 230 V, 6-steps
50, 15A
415/0-470V, 15A
QTY.
1 no.
1 no.
1 no.
1no.
1 no
1 set
CIRCUIT DIAGRAM:
THEORY :
The basic principle used for measuring active as well as reactive power in an a.c. circuit is the
Blondels theorem.
If a network is supplied through n conductors, the total power is measured by summing the
readings of n wattmeters is in each line and the corresponding voltage element is connected between the
line and a common point.
If the common point is located on one of the lines, then the power may be measured by (n-1)
wattmeters.
One wattmeter method can be used for the measurement of power but this method is applicable
only for balanced loads the current coil is connected in one of the lines and one end of the pressure coil to
some line, other end being connected alternatively to other two lines.
In case of balanced three phase circuits it is simple to use a single wattmeter to read reactive
power the current coil of the wattmeter is connected in one line and pressure coil is connected across other
two lines.
Reading of wattmeter
= V13 I 2 cos(90 + )
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
3 VI cos(90 + )
3 VI sin
=
Q = 3 VI sin = 3 x reading of wattmeter.
1 Q
Phase angle = tan .
P
PROCEDURE :
1.
2.
3.
4.
5.
6.
TABULAR FORM:
S.N
o.
Voltmeter
reading
V(volts)
Ammeter
reading
I(Amps)
Wattmeter
reading W
(watts)
sin= Q/ 3 VI
PRECAUTIONS :
1.
Keep the 3-phase variac in zero position initially.
2.
Keep inductive load at no-load position initially.
3.
Live terminals should not be touched.
4.
Connect the circuit without any loose connection.
5.
If any wattmeter reads, reading, change either current coil or pressure coil connections.
6.
Take all meter readings without parallel error.
7.
Load current should not exceed rated current value.
8
Vary the variac such that the current and voltage are within the rated value
9.
Meter readings should not exceed their ratings.
10
Load should be varied very smoothly.
11
Take proper care, such that the ammeter reading, should not exceed the current rating of LPF
wattmeter.
RESULT : Reactive power absorbed by the inductive load is calculated & tabulated for various loading
conditions.
QUESTIONS:
1.
2.
3.
4.
5.
6.
7.
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
EXPERIMENT NO.8
MEASUREMENT OF PARAMETERS OF CHOKE COIL USING 3VOLTMETERS &
3AMMETERS METHOD
AIM : To obtain the parameters (R, X, L,Z, power and PF)of given choke coil using 3 Voltmeter and 3
Ammeter methods.
APPARATUS
Sl. NO.
1.
2.
3.
4.
5.
6
7
:
NAME
Choke coil
.Regulated D.C
Power supply
Voltmeter
Ammeter
Rheostat
Single Phase Variac
Connecting wires
TYPE
RANGE
230 V, 10 A
(0-30V)
QTY.
1 no.
1 no.
MI
MI
300 V
10A
50, 15A
0-270V, 10A
3 no.
4 no.
1no.
1 no
1 set
AC
CIRCUIT DIAGRAM:
THEORY :
An inductive transducer works on the principle of variation of inductance using multiple coils.
The coils that are being used need to be evaluated and their parameters so defined such that the use of their
parameters may be regarded as constant and accurate.
Thus emphasis need to be laid upon the method of measurement of inductance of choke coil by using 3
voltmeter meter method and 3 ammeter method.
For 3-Voltmeter method:
V12 = V22 + V32 2V2V3 cos
cos =
r
V12 V22 V32
, but cos = 2
[
r
+
(
2
fL) 2 ]1 / 2
(2V2V3 )
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
2
(2V2V3 )
r + (2fL ) 2
r + (2fL ) = r
(2fL) 2 =
L=
1
2f
V22
1/ 2
4V22V32
4V22V32 r 2
V12
r2
2
V32
4V22V32r 2
2
V12 V22 V32
r2
I 12 = I 22 + I 23 2I 2I 3 cos
cos =
r
I 12 I 22 I 32
; 2
(2 I 2 I 3 )
r + (2fL ) 2
I12 I 22 I 32
r
=
2
2 I 2 I3
r + (2fL ) 2
L=
1
2f
4 r 2 I 22 I 32
I 12
I 22
2
I 32
1/ 2
1/ 2
r2
Hence by using the above formulae we can calculate the inductance of a choke coil.
PROCEDURE :
3 Voltmeter method
1.
Connections are made as per the circuit diagram.
2.
Keep the variac in minimum position and close the DPST switch.
3.
Vary the variac gradually in steps up to rated current capacity of conductor that used for coil or up
to voltage less than rated voltage indicated by voltmeter1
4.
Note down the readings of meters in each step. Tabulate the readings as per tabular form1
5.
Gradually vary the variac to minimum position and open the DPST switch.
6
Now calculate the parameters of the choke coil.
3 Ammeter method
1.
Connections are made as per the circuit diagram.
2.
Keep the variac in minimum position and close the DPST switch.
3.
Vary the variac gradually in steps upto the rated voltage (230V).OR Apply current less than
(500mA) indicated by ammeter 2
4.
Note down the readings of 3 Ammeters in each step. Tabulate the readings as per tabular form 2
5.
Gradually change the variac to minimum position and open the DPST switch.
6.
Now calculate the parameters of the choke coil.
MEASUREMENTS OF RESISTANCE OF CHOKE COIL:
1.
2.
3.
4
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
Calculate resistance values R10v, R15v, R20v and take average of 3 resistance values as resistance
of choke coil.
TABLE 1
S.NO
V1
TABLE 2
S.NO
V
V2
V3
IL
coso
ZL
RL
XL
I1
I2
I3
COSO
ZL
RL
XL
PRECAUTIONS :
1
2
3
4
5
6
7
8
9
10
11
RESULT :
The experiment is conducted by 3-voltmeter and 3-ammeter methods and inductance of the given choke
coil is calculated
QUESTIONS:
1.
2.
3.
4.
5.
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
Calculation:
At on load
i.e., IL = 1 amp
V2 = 100 v
V3 = 106 v
P = 230 watts
Cos = 0.9
Therefore V12 = V22 +V32- 2 V2 V3 cos = 220 v
L = (1/2f) * ((4 V22 V32 r2)/ (V12 -V22 -V32)) - r2
Here f = 50Hz
L =(1/2f) * ((4 I22 I32 r2)/ (I12 -I22 -I32)) - r2
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
EXPERIMENT NO.9
CALIBRATION OF LPF WATTMETER BY PHANTOM LOADING
AIM : Testing and Calibration of the given LPF wattmeter by using a sub-standard Wattmeter by phantom
loading method.
APPARATUS :
Sl. NO.
1
2.
3.
4.
5.
6
7
NAME
Wattmeter(Testing or standard)
Wattmeter(Sub-standard or calibrated)
Voltmeter
Ammeter
Rheostat
Single Phase Variac
Connecting wires
TYPE
Dynamometer type, LPF
Dynamometer type, UPF
MI
MI
RANGE
10A, 300V
10A, 300V
300 V
10A
350, 1.2A
0-270V, 10A
QTY.
1 no
1 no.
1 no.
1 no.
1no.
1 no
1 set
CIRCUIT DIAGRAM:
THEORY :
Electrodynamometer type wattmeter has two coils connected in different circuits for measurement
of power. The fixed coils or field coils are connected in series with the load and so carry the current in
the circuit. The fixed coils, therefore, form the current coil or simply C.C of the wattmeter. The
moving coil is connected across the voltage and, therefore, carries a current proportional to the voltage. A
high non-inductive resistance is connected in series with the moving coil to limit the current to a small
value. Since the moving coil carries a current proportional to the voltage, it is called the pressure coil or
voltage coil or simply called P.C. of the wattmeter.
Both fixed and moving coils are air cored. The voltage rating of the wattmeter is limited to about 600 V by
the power requirements of the voltage circuit since most of the power is absorbed by the resistance in series
with the moving coil and considerable heat is generated. For higher voltages, the pressure coil circuit is
designed for 110V, and a potential transformer is used to step down the voltage.
If V= Voltage across the pressure coil,
I = Current through the current coil and
= Angle between V& I, then
P = Power being measured = VI cos
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
The deflection is directly proportional to the power being measured and the scale is essentially uniform
over the range in which (dM/d ) is constant. By suitable design, the mutual inductance between fixed
and moving coil and be made to vary linearly with angle over a range of 40o to 50o on either side of zero
mutual inductance position.
On lagging loads the wattmeter will read high, as the effect of the inductance of the pressure coil circuit is
to bring the pressure coil current more nearly into phase with the load current than would be the case if this
inductance were zero. Very serious errors may be introduced by pressure coil inductance at low power
factors unless special precautions are taken.
Many wattmeters are compensated for errors caused by inductance of pressure coil by means of a capacitor
connected in parallel with a portion of multiplier (series resistance).
Measurement of power in circuits having low power factor by ordinary electrodynamometer wattmeter is
difficult and inaccurate because:
i) The deflecting torque on the moving system is small (owing to low power factor) even when
the current and pressure coils are flly excited;
ii) Errors introduced because of inductance of pressure coil tend to be large at low power factors.
Special features are incorporated in an electrodynamometer wattmeter to make it a low power factor type of
wattmeter. These features are discussed in details below:
1. Pressure Coil Current. The pressure coil circuit is designed to have a low value of resistance,
so that the current, flowing through it, is increased to give an increased operating torque. The pressure coil
current in a low power factor wattmeter may be as much as 10 times the value employed for high power
factor wattmeter.
2. Compensation for Pressure coil Current. The power being measured in a low power factor
circuit is small and current is high on account of low power factor. It is absolutely necessary to compensate
for the pressure coil current in low power factor wattmeter.
When the current rating of meter under test is high a test with actual loading arrangements would involve a
considerable waste of power. In order to avoid this Phantom loading as Fictitious loading is done. Phantom
loading consists of supplying the pressure coil from a circuit of required normal voltage and the current coil
from a separate low voltage supply. It is possible to circulate the rated current through the current coil with
a low voltage supply as the impedance of this circuit is very low, with this arrangement. The total power
supplied for the test is that due to the small pressure coil current at normal voltage, plus that due to the
current circuit current applied at low voltage. The total power therefore, required for testing the meter with
phantom loading is comparatively small.
PROCEDURE :
1.
Connections are made as per the circuit diagram.
2.
With variac I and II in minimum position and the rheostat in minimum position, close the
DPST-I switch to connect the supply-I.
3.
Vary the variac-I gradually to apply the rated voltage (230V). to pressure coils of both p.f meter
and wattmeter
4
Close the DPST-II switch and vary the variac-II to adjust a suitable current of (1A or 2A) in the
current coils of LPF wattmeter and p.f meter. as indicated by ammeter. Note down the readings
of all the instruments
5
Now vary the current (upto 4A) in current coils of both wattmeter and p.f meter in steps using
single phase variac & rheostat and note down readings for each step. Tabulate the readings.
6
Bring the rheostat to minimum position, vary the variac-II and variac-I to minimum position and
open the DPST switches.
7.
Evaluate True power (S1) , Actual power (S2) , %error, %correction.
www.jntuworld.com
www.jntuworld.com
8.
www.jwjobs.net
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
TABULAR FORM:
S.No
Voltmeter
Reading
V (volts)
Ammeter
Reading,
I (Amps)
P.F
meter
reading
Wattmeter
Reading(S2)
(Watts)
% error
(S1-S2)/S1
MODEL CALCULATIONS:
True power (S1)
Actual power (S2)
% error
%correction
= V * I *cos
= Wattmeter reading.lpf
= [(S1- S2) / S1] * 100
= [(S2 S1) /S1] * 100
MODEL GRAPH :
%error
Load current
PRECAUTIONS:
1.
2.
3
4
5
6
7
9
10
11
12
RESULT:
The LPF wattmeter is calibrated using phantom loading and calibration curve is plotted
for the given wattmeter
QUESTIONS:
1.
2.
3.
4.
5.
6.
7.
8.
9.
www.jntuworld.com
%Correction
(S2-S1)/S1
www.jntuworld.com
www.jwjobs.net
Calculations:
True power S1 = VICos
Actual power S2 = wattmeter reading LPF
% error = ((S1-S2)/S1)*100
%correction = ((S2-S1)/S1)*100
By taking values
V = 230 v
I = 1 amp
S1 = 230 * 1* 0.95 =218.5 watts
S2 = 90 watts
% error = ((S1-S2)/S1)*100 = 58.81%
%correction = ((S2-S1)/S1)*100 = -58.81%
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
EXPERIMENT NO.10
MEASUREMENT OF 3-PHASE ACTIVE POWER BY USING 2 CTS AND SINGLE WATTMETER
METHOD
AIM : To measure the 3-phase active power by using a single wattmeter and 2 CTs.
APPARATUS :
Sl. NO.
1
2.
3.
4.
5.
6
7
NAME
Wattmeter
CTs
Voltmeter
Ammeter
3-phase variable inductive load
Three Phase Variac
Connecting wires
TYPE
Dynamometer type, LPF
MI
MI
Balanced star
RANGE
10A, 300V
1:1, 5/5A
300 V
10A
350, 1.2A
0-270V, 10A
QTY.
1 no
2 no.
1 no.
2 no.
1no.
1 no
1 set
CIRCUIT DIAGRAM
THEORY :
To measure 3- phase power using two CTs. Power can be measured in many ways, for an balanced load,
only one wattmeter is enough to measure the 3- phase power, and for an unbalanced load, two wattmeter
method is used to measure the 3- phase power. The primary windings of CTs are connected in series with
two phases. The secondary of both the CTs are connected as shown in figure the current coil of wattmeter
connected across both secondaries of CTs. The pressure coil is connected between the two phases.
PROCEDURE :
1.
Make the connections as per the circuit diagram.
2.
Keep the 3 phase variac in minimum position and close the TPST switch.
3.
Vary the 3- phase variac gradually and apply the rated voltage (415V).
4.
Note down all the meter readings and tabulate them.
5.
Vary the inductive load in steps upto the rated current and tabulate the meter readings in each step.
6.
Reduce the voltage to zero gradually by varying the 3 variac and open the TPST switch.
TABULATION:
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
S.NO
Ict1
Ict2
load
PRECAUTIONS :
1.
Avoid loose connections.
2.
The secondary of CT should not be kept open.
3.
Take readings without error.
4.
Keep variac at their minimum position initially
5.
Vary the variac such that the current and voltage are within the rated value
6.
Meter readings should not exceed their ratings.
7.
Live terminals should not be touched.
8
If any wattmeter reads zero or negative reading, change either current coil or pressure coil
connections.
9.
Load current should not exceed rated current value.
10.
Load should be varied very smoothly.
13.
Take proper care, such that the ammeter reading, should not exceed the current rating of LPF
wattmeter.
RESULT :
The 3- active power of the inductive load is calculated using single phase wattmeter and two CTs
method. at balanced load condition.
QUESTIONS:
1.
2.
3
4
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
EXPERIMENT NO.11
LVDT AND CAPACITANCE PICKUP-CHARACTERISTICS AND CALIBRATION
AIM:
To measure the displacement using linear variable differential transformer.
APPARATUS:.
CIRCUIT DIAGRAM:
THEORY:
Linear variable differential transformer LVDT is a transducer. Basically it is passive inductive transformer
similar to a potential transformer.
LVDT consists of three windings, one primary and two secondaries of equal turns. Primary is wound
centrally between two secondaries. All three windings are wound on a hollow tubular former through
which magnetic core slides.
Core affects magnetic coupling between primary and the secondaries while primary is connected to an AC
signal.
Normal / null position of core causes equal induced voltage in both the secondaries. Hence the total
difference voltage of both the secondaries becomes zero. Any deviation in core position from its null
position induces unequal voltage from both secondaries and hence the difference signal of it is a non zero
quantity, this non zero quantity varies with core position. Ideally displacement versus change in difference
signal should be linear.
When ES1=ES2 (core at null position or central position)
Ediff=0
When core is moved left
ES1>ES2 &
Ediff (ES1-ES2) is in phase with ES1
When core is moved right ES1<ES2
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
Observations:
GRAPH:
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
PRECAUTIONS:
1.
2.
Allow the instrument in ON position for 10 min. for initial warm up.
While calibrating vary the cal/span pot slowly and smoothly.
RESULT:
Displacement is measured using linear variable differential transformer and graphs of indicated
displacement Vs actual displacement, %Error Vs actual displacement and output voltage Vs actual
displacement.
QUESTIONS:
1.
2.
3.
4.
5.
What is LVDT?
What is transducer?
How many transducers are there?
How many windings the transformer in LVDT have in its construction?
How the secondaries are connected in the transformer of LVDT?
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
EXPERIMENT NO.12
MEASUREMENT OF IRON LOSS IN A BAR SPECIMEN USING A CRO AND USING A
WATTMETER
AIM:
To measure the iron losses in strip (sheet) material using Lloyd - Fisher Square.
APPARATUS:
CIRCUIT DIAGRAM
THEORY:
The voltage across the secondary i.e. voltmeter reading given the rms value of the induced voltage.
E = 4 kfm f N2
N2 = No. of turn of the secondary winding.
E
Bm1 = AK f AsfN 2 (apparent value)
As = area of specimen; m2
Bm = Bm1 0Hm
Ac
1
As
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
Wattmeter reading = iron loss in the specimen + Copper loss in the secondary
E2
Cu loss = rp + rc
Where rp = resistance of pressure coil
Rc = resistance of coil S1.
PE
E2
3P
Eddy current loss = Volume x
3.
4.
5.
OBSERVATIONS :
CALCULATIONS:
P = Wattmeter reading
P
Pi = iron loss =
V2 reading
V22
V1 reading rp + rc
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
PRECAUTIONS:
1.
2.
3.
type.
Take the meter readings without parallel error.
www.jntuworld.com
www.jntuworld.com
www.jwjobs.net
RESULT:
Iron loss are calculated in a bar specimen by using Lloyed fisher square method, by conducting the
experiment.
QUESTIONS:
1.
2.
3.
4.
5.
6.
7.
8.
9.
www.jntuworld.com