Calibration
Calibration
CALIBRATION
CALIBRATION
CALIBRATION
Procedures,
Procedures,
Standard
Standard
And
And
Methodologies
Methodologies
Introduction
In this section, the procedures and
methodologies to carry out a
calibration will be discussed. In
addition, the approach to calculate
errors in a calibration will be shown.
Objective Of Calibration
The prime objective of calibration is
to ensure that measurements taken
are accurate at all times taking into
consideration wear and tear,
degradation and abuse. It is with this
in mind that we need to adopt and
appropriate methodology to do a
proper calibration.
cope of Calibration
In developing a calibration program, the following
should be taken into consideration !
If you are unsure which is the best interval to choose, always follow the e"uipment
manufacturer<s recommendations. This interval should be reviewed from time and as a
rule of thumb, it should be adjusted based on the calibration history. -or e(ample, if
an e"uipment is calibrated on a si( monthly basis and it is found to be out of
calibration fre"uently, then the calibration interval should be changed either 5 or =
monthly.
C'9I>)'TIO& I&T#)8'9
*hat actions and the calibration interval on the e"uipment would you recommend to
the user for the following !
?. 'n instrument is found to have failed 5 times in the last 5 calibrations
performed at @ monthly intervals.
5. ' newly purchased p$ onAline analy.er which will be used in production daily.
Calibration interval recommended by manufacturer is = monthly.
=. ' +agnetic -low +eter used to measure the flow of water with
specification of 3.6 to ? B accuracy.
'ccuracy of +agnetic -low +eter is 3.6 to ?B with Calibration interval of
yearly using the -low +eter 8erificator.
'CC/)'CC O- )#-#)#&C# T'&7')7
's discussed in the previous section, the
reference standard used to do a calibration must
be at least D to ?3 times more accurate than the
e"uipment. In general, a 26B confidence level is
preferred and this gives a E 5 sigma (standard
deviation) or D sigma capability in a measurement.
O/T O- TO9#)#&C# #F/I%+#&T
*hen an e"uipment is found to be out of tolerance or not having the re"uired
accuracy, certain steps must be taken. 'ctions to be taken include decisions on
whether the e"uipment can be used or not and what needs to be done on the affected
products as follows !
; )eview all measurements prior to and after adjustment is made to the
affected e"uipment.
; 7ecide on whether the e"uipment can be used or has to be thrown away
or used for lesser accuracy measurements.
; )eview previous batches of products 0 ? week or more1 where
measurements have been taken using this out of tolerance e"uipment
prior to and after an adjustment has been made to correct the out of
calibration situation. 7ecide whether they meet specification or not.
; If the e"uipment can be used but needs correction to their readings, then the
compensation or offset factors must be stated on the e"uipment.
ubcontractor Controls
In choosing an appropriate subcontractor to perform your
calibration, it is critical that an accredited calibrating lab
or agent is used 0those certified to IO 2333 4uide 561.
'sk for a copy of their certification before selecting them.
Cou will also need to ensure that the reference standards
used by the calibrating lab or agent is traceable to national
standard. The calibration method used should also be
checked to ensure they used an appropriate or recogni.ed
procedure. >oth of these re"uirement can be verified on
their Certificate of Calibration, which is issued to you upon
completion of a calibration for each set of e"uipment.
/ncertainty Of +easurement
The uncertainty of measurement of a calibration needs to be
calculated to determine all the possible sources of errors. It is
given by the formula !
/ncertainty of +easurement G E t 0from a tudents tAtable1 ( s
0standard deviation of variation1.
As a rule of thumb, the uncertainty of measurement must not exceed
30% of the tolerance of a measurement. If it does, its mean that the
equipment is not suitable for use as its required accuracy is not
sufficient. A measurement taken using this equipment ill gi!e a false
or inaccurate !alue.
$andling and torage
The proper handling and storage of measuring e"uipment and their
reference standards are an essential part of establishing an
effective Calibration %rogram. ome of the guidelines to follow are
as follows!
$andling : ensure the e"uipment is appropriately handled during use, storage and
transportation.
Check that the calibration status is valid before issuing an e"uipment for use
0especially newly purchased e"uipment1.
9ocation
)e"uired 'ccuracy
+easurement )ange
Calibration Interval
7ate Calibrated
)ecording methodologies K
Calculations K
date of purchase