Introduction To CT Theory
Introduction To CT Theory
Chapter 1
Introduction to
CT physics
Image generation
Since the first CT scanner was developed in 1972 by Sir Godfrey Hounsfield,
the modality has become established as an essential radiological technique
applicable in a wide range of clinical situations.
CT uses X-rays to generate cross-sectional, two-dimensional images of the
body. Images are acquired by rapid rotation of the X-ray tube 360° around the
patient. The transmitted radiation is then measured by a ring of sensitive radiation
detectors located on the gantry around the patient (Fig. 1.1). The final image is
generated from these measurements utilizing the basic principle that the internal
structure of the body can be reconstructed from multiple X-ray projections.
Early CT scanners acquired images a single slice at a time (sequential
scanning). However, during the 1980s significant advancements in technology
heralded the development of slip ring technology, which enabled the X-ray tube
to rotate continuously in one direction around the patient. This has contributed
to the development of helical or spiral CT.
In spiral CT the X-ray tube rotates continuously in one direction whilst the
table on which the patient is lying is mechanically moved through the X-ray
beam. The transmitted radiation thus takes on the form of a helix or spiral. Instead
of acquiring data one slice at a time, information can be acquired as a continuous
volume of contiguous slices (Fig. 1.2a, b). This allows larger anatomical regions
of the body to be imaged during a single breath hold, thereby reducing the
possibility of artefacts caused by patient movement. Faster scanning also increases
patient throughput and increases the probability of a diagnostically useful scan
in patients who are unable to fully cooperate with the investigation.
3
Ch01.qxd 7/5/04 10:06 AM Page 4
Patient
4
Ch01.qxd 7/5/04 10:06 AM Page 5
Patient/table
movement
Fig. 1.2 (A) Single-slice system (one ring). (B) Single-slice helical CT. The
X-ray tube rotates continuously and the patient moves through the X-ray beam
at a constant rate.
5
Ch01.qxd 7/5/04 10:06 AM Page 6
Patient/table
movement
Fig. 1.3 (A) Multidetector system (four rings shown here). (B) Multislice
helical CT.
6
Ch01.qxd 7/5/04 10:06 AM Page 7
Each pixel is assigned a numerical value (CT number), which is the average of all
the attenuation values contained within the corresponding voxel. This number is
compared to the attenuation value of water and displayed on a scale of arbitrary
units named Hounsfield units (HU) after Sir Godfrey Hounsfield.
This scale assigns water as an attenuation value (HU) of zero. The range of CT
numbers is 2000 HU wide although some modern scanners have a greater range
of HU up to 4000. Each number represents a shade of grey with +1000 (white)
and –1000 (black) at either end of the spectrum (Fig. 1.4).
Air Water
7
Ch01.qxd 7/5/04 10:06 AM Page 8
Whilst the range of CT numbers recognized by the computer is 2000, the human
eye cannot accurately distinguish between 2000 different shades of grey.
Therefore to allow the observer to interpret the image, only a limited number of
HU are displayed. A clinically useful grey scale is achieved by setting the WL and
WW on the computer console to a suitable range of Hounsfield units, depending
on the tissue being studied.
The term ‘window level’ represents the central Hounsfield unit of all the
numbers within the window width.
The window width covers the HU of all the tissues of interest and these are
displayed as various shades of grey. Tissues with CT numbers outside this range
are displayed as either black or white. Both the WL and WW can be set
independently on the computer console and their respective settings affect the
final displayed image.
For example, when performing a CT examination of the chest, a WW of 350
and WL of +40 are chosen to image the mediastinum (soft tissue) (Fig. 1.5a),
whilst an optimal WW of 1500 and WL of –600 are used to assess the lung fields
(mostly air) (Fig. 1.5b).
What is pitch?
Pitch is the distance in millimetres that the table moves during one complete
rotation of the X-ray tube, divided by the slice thickness (millimetres). Increasing
the pitch by increasing the table speed reduces dose and scanning time, but at the
cost of decreased image resolution (Fig. 1.6a, b).
Image reconstruction
The acquisition of volumetric data using spiral CT means that the images can be
postprocessed in ways appropriate to the clinical situation.
● Multiplanar reformatting (MPR) – by taking a section through the three-
dimensional array of CT numbers acquired with a series of contiguous slices,
sagittal, coronal and oblique planes can be viewed along with the standard
transaxial plane (Fig. 1.7).
8
Ch01.qxd 7/5/04 10:06 AM Page 9
Fig. 1.5 These two images are of the same section, viewed at different
window settings. (A) A window level of +40 with a window width of 350
reveals structures within the mediastinum but no lung parenchyma can be seen.
(B) The window level is –600 with a window width of 1500 Hounsfield units.
This enables details of the lung parenchyma to be seen, at the expense of the
mediastinum.
9
Ch01.qxd 7/5/04 10:06 AM Page 10
Fig. 1.6 (A) Pitch is low. The table moves less for each tube revolution. The
image is sharper. (B) Pitch is high. The table moves further for each revolution
so the resulting image is more blurred. The helix is stretched.
10
Ch01.qxd 7/5/04 10:06 AM Page 11
B C
11
Ch01.qxd 7/5/04 10:07 AM Page 12
A B
Contrast media
Contrast between the tissues of the body can be improved by the use of various
contrast media. These mostly contain substances with a high molecular weight
and thus increase the attenuation value of the organ they opacify.
12
Ch01.qxd 7/5/04 10:07 AM Page 13
Contrast media
Dual-modality imaging
13
Ch01.qxd 7/5/04 10:07 AM Page 14
14
Ch01.qxd 7/5/04 10:07 AM Page 15
Fig. 1.9 cont’d. (C) PET/CT fused image correlating increased metabolic
activity within lymph node masses implying pathological enlargement.
Courtesy: Nuclear Medicine, University Hospital, Zurich, Switzerland.
Advantages
● CT is readily available in most hospitals.
● It is an increasingly rapid imaging modality with excellent image resolution,
enabling faster and more accurate diagnostic evaluation of patients over a
wide spectrum of clinical indications.
● The data acquired in one scan can subsequently be manipulated to provide
multiplanar and 3D reconstructions.
Disadvantages
● Radiation – although CT scans account for only 4% of X-ray examinations,
they contribute to more than 20% of the radiation dose to the population by
‘medical X-rays’. For typical doses of common radiological examinations, see
Table 1.1.
15
Ch01.qxd 7/5/04 10:07 AM Page 16
Approx. equiv.
Diagnostic Typical effective Equiv. no. period of
procedure dose (mSv) of CXR background radiation
UK average background radiation = 2.2 mSv per year; regional averages range from
1.5 to 7.5 mSv per year.
16