Preventive Maintenance Preventive Maintenance Checklist
UV/VIS Excellence
Company information
Depertment of Environment and Natural Rersources Instrument location: Inst. Room
East Avenue Diliman Quezon City
1101 Philippines Instrument asset number: SPC-03
Device information
Model: UV5 Instrument serial number: C032807645
Instrument FW version: 3.0.1
Spectrograph serial number: C031774080
Instrument status - As found
Report date and time
Data backup Passed
Performance test Oct 21, 2024 Passed
Results and history analysis Passed
Visual checks
External power supply adapter Passed
Housing, surfaces and covers Passed
Screws Passed
Rubber feet Passed
Terminal including cable Passed
Filter mat Passed
Display protective cover Passed
FillPalMini/SPR200 N/A
Cuvette holder Needs attention
CuvetteChanger N/A
Cleaning
Terminal Performed
Housing, surfaces and covers Performed
Ball lenses Performed
FillPalMini/SPR200 N/A
Cuvette holder Performed
CuvetteChanger N/A
METTLER TOLEDO Service ME-30652007 v2.2 1/2
Version date: August 2021
UV/VIS Preventive Maintenance Checklist
Operational checks
Report date and time
Terminal check Passed
On/Off switch Passed
Main board tests Passed
Preventive Maintenance
Device board test Passed
Fan and temperature sensor check Passed
A/D converter and lamp adjustment N/A
CuvetteChanger adjustment (position) N/A
Instrument status - As left
Report date and time
Performance test Oct 21, 2024 Passed
Data backup provided to device owner Performed
Results discussed with device owner Performed
The UV/VIS Excellence Spectrophotometer listed in this document has been cleaned and checked according to the METTLER TOLEDO
Preventive Maintenance SOP ME-30257265.
Note that evidence of the results generated by the instrument is attached. The preventive maintenance was carried out and the results
were obtained under the prevailing ambient conditions.
Technician information
Remarks & recommendations
Unit operational.
Service provider name and address: Mettler Toledo Philippines Muntinlupa City
Performed by: Jovil Dela Cruz Date: Oct 21, 2024
Signature:
METTLER TOLEDO Service ME-30652007 v2.2 2/2
Version date: August 2021