0% found this document useful (0 votes)
30 views2 pages

UV/VIS Preventive Maintenance Report

The document outlines the preventive maintenance checklist for the UV/VIS Excellence Spectrophotometer, detailing the instrument's status, visual checks, cleaning, and operational checks. All checks were passed, except for the cuvette holder, which needs attention. The maintenance was performed by Jovil Dela Cruz on October 21, 2024, and results were discussed with the device owner.

Uploaded by

Eddard Stark
Copyright
© All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
30 views2 pages

UV/VIS Preventive Maintenance Report

The document outlines the preventive maintenance checklist for the UV/VIS Excellence Spectrophotometer, detailing the instrument's status, visual checks, cleaning, and operational checks. All checks were passed, except for the cuvette holder, which needs attention. The maintenance was performed by Jovil Dela Cruz on October 21, 2024, and results were discussed with the device owner.

Uploaded by

Eddard Stark
Copyright
© All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Preventive Maintenance Preventive Maintenance Checklist

UV/VIS Excellence

Company information

Depertment of Environment and Natural Rersources Instrument location: Inst. Room


East Avenue Diliman Quezon City
1101 Philippines Instrument asset number: SPC-03

Device information
Model: UV5 Instrument serial number: C032807645

Instrument FW version: 3.0.1

Spectrograph serial number: C031774080

Instrument status - As found


Report date and time

Data backup Passed

Performance test Oct 21, 2024 Passed

Results and history analysis Passed

Visual checks

External power supply adapter Passed

Housing, surfaces and covers Passed

Screws Passed

Rubber feet Passed

Terminal including cable Passed

Filter mat Passed

Display protective cover Passed

FillPalMini/SPR200 N/A

Cuvette holder Needs attention

CuvetteChanger N/A

Cleaning

Terminal Performed

Housing, surfaces and covers Performed

Ball lenses Performed

FillPalMini/SPR200 N/A

Cuvette holder Performed

CuvetteChanger N/A

METTLER TOLEDO Service ME-30652007 v2.2 1/2


Version date: August 2021
UV/VIS Preventive Maintenance Checklist

Operational checks
Report date and time

Terminal check Passed

On/Off switch Passed

Main board tests Passed


Preventive Maintenance

Device board test Passed

Fan and temperature sensor check Passed

A/D converter and lamp adjustment N/A

CuvetteChanger adjustment (position) N/A

Instrument status - As left


Report date and time

Performance test Oct 21, 2024 Passed

Data backup provided to device owner Performed

Results discussed with device owner Performed

The UV/VIS Excellence Spectrophotometer listed in this document has been cleaned and checked according to the METTLER TOLEDO
Preventive Maintenance SOP ME-30257265.

Note that evidence of the results generated by the instrument is attached. The preventive maintenance was carried out and the results
were obtained under the prevailing ambient conditions.

Technician information
Remarks & recommendations

Unit operational.

Service provider name and address: Mettler Toledo Philippines Muntinlupa City

Performed by: Jovil Dela Cruz Date: Oct 21, 2024

Signature:

METTLER TOLEDO Service ME-30652007 v2.2 2/2


Version date: August 2021