Lab 5

The document outlines the objectives and procedures for a lab focused on gantry motion and patient table adjustments, including calibration and tension checks. It emphasizes the proprietary nature of the content, restrictions on distribution, and includes safety precautions for handling equipment. The lab is designed for Philips personnel and licensed representatives, requiring specific tools and reference materials for completion.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
69 views7 pages

Lab 5

The document outlines the objectives and procedures for a lab focused on gantry motion and patient table adjustments, including calibration and tension checks. It emphasizes the proprietary nature of the content, restrictions on distribution, and includes safety precautions for handling equipment. The lab is designed for Philips personnel and licensed representatives, requiring specific tools and reference materials for completion.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

CTI2IL016 Gantry Motion and Patient Table Lab

Proprietary Notice:
This document and the information contained in it is proprietary and confidential information of Philips and
may not be reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disseminated
without the prior written permission of the Philips Legal Department. Use of this document and the
information contained in it is strictly reserved for current Philips personnel and Philips customers who have a
current and valid license from Philips for use by the customer’s designated in-house service employee on
equipment located at the customer’s designated site. Use of this document by unauthorized persons is strictly
prohibited. Report violation of these requirements to the Philips Legal Department. This document must be
returned to Philips when the user is no longer licensed and in any event upon Philips’ first written request.

CSIP Level 2:
This document or digital media and the information contained in it is restricted proprietary and confidential
information that CANNOT BE DISTRIBUTED OR LICENSED AND IS ONLY FOR THE USE OF PHILIPS PERSONNEL
AND LICENSED REPRESENTATIVES. IT IS NOT TO BE DISTRIBUTED, SHARED, COPIED, TRANSFERRED OR
OTHERWISE PROVIDED TO ANYONE ELSE, ELECTRONICALLY OR OTHERWISE.

Warranty Disclaimer Language:


Philips provides this DOCUMENT without warranty of any kind, implied or expressed, including, but not
limited to, the implied warranties of merchantability and fitness for a particular purpose.

Limitations of Liability Language:


Philips has taken care to ensure the accuracy of this document. However, Philips assumes no liability for
errors or omissions and reserves the right to make changes without further notice to any products herein to
improve reliability, function, or design. Philips may make improvements or changes in the product(s) or
program(s) described in this document at any time.

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 1 of 7
Objectives
When you have successfully completed this lab, you will be able to:
• Check and Adjust the Couch
• Check and Adjust the Tilt mechanism (Not for Access CT)
• Configure the main drive servo controller
• Configure the patient table horizontal servo controller
• Check the balance of the gantry
• Check the tension of the belts

Prerequisite Lessons
• Gantry Motion Lecture
• Patient Table Lecture

Tools/Reference Materials
• Lab Toolkit

Time
2.5 hours

Introduction
This lab covers the rotor motion, patient table horizontal motion and if applicable tilt motion and patient table
vertical motion

Procedure Preview
In this lab exercise, the learner will perform
• Rotor Calibration
• Couch Vertical Calibration
• Couch Horizontal Calibration
• Tilt Calibration (MX16/Incisive CT)
• Measurement of the Belt tension

Special Instructions/Precautions
Make sure you read and understand the safety guidelines before proceed with the lab assignments

<CAUTION:
Damage to the equipment and personnel can occur if you are not careful. If you do not know the outcome of
a motion before initiating, DON’T DO IT! Please seek help from the instructor if unsure.>

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 2 of 7
Instructions
1. Couch Limit Switched Recognition
a. Look for all micro switches related to horizontal motion.
b. List all the micro switches and their functionalities

Micro Switch Function


SL301 Horizontal In Limit Switch

c. Look for all micro switches related to vertical motion


d. List all the micro switches and their functionalities

Micro Switch Function


SL305 Vertical Down Position Switch

2. Lifted Patient Table Vertical Adjustment


a. Perform the vertical adjustment as descripted in the Calibration and Adjustment Instructions Manual.
b. Only check the principle of the “Down Limit Switch” SL305. DO NOT TOUCH. This switch is checked
and calibrated in the factory. Changing the position might cause table error ERR03.

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 3 of 7
3. In the following exercise you will go through the steps that need to be done when servicing
the Gantry motion parts.
a. Which picture shows positive tilt?

A B

b. Use the calibration manual to calibrate the tilt.


c. Determine the tilt envelope using the table
Table Height [mm]
25 30 35 40 45 50 55 60
0 0 0 0 0 0 0 0

Positive Tilt
angle

Negative Tilt
angle

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 4 of 7
d. Use the graphic to draw the envelope
300 350 400 450 500 550 600
30

20

10

-10

-20

-30

e. Use the calibration manual to check the Tilt angle range adjustment

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 5 of 7
4. Rotor Belt Tension Adjustment
a. Perform the belt tension adjustment as descripted in the Calibration and Adjustment Instructions
Manual.

5. Servo Controller Programming


a. Program the main drive servo controller as descripted in the CT System Flash Programming Manual
b. Program the patient table horizontal servo controller as descripted in the CT System Flash
Programming Manual

6. Service Tools (Auto Calibration) – Gantry Balance Check


a. Go to Service Tools
b. At “Advance Tab”, Select “Auto Calibration”
c. At “Calibration Process” column, Select “Custom Calibration”
d. Run Rotor Balance Test

—End of Lab—

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 6 of 7
Content Change Summary

Revision Date Modified By Revision Summary


(dd-MMM-yyyy) (Name)
06-MAR-2019 Frank Broeders Initial Release

Template Change Summary

Revision Date Modified By Revision Summary Approver Name


(yyyy-MMM-dd) (Name) & Date (yyyy-MMM-DD)
Initial Release
2015-JAN-01 D. Preston Changed copyright year to 2015 G. Argento 2015-JAN-01
2015-SEP-18 G. Argento Changed this column’s title G. Argento 2015-SEP-18
2016-JAN-06 D. Preston Changed copyright year to G. Argento 2016-JAN-06
2016
2016-NOV-23 G. Argento Changed copyright year to 2017 G. Argento 2016-NOV-23
2018-MAY-09 D. Preston Changed copyright year to 2018. Sue Hove 2018-May-10
To comply with GCS-ED-PR-00003:
Added “Content Change Summary”
table; changed this table to
“Template Change Summary”;
changed date format to YYYY-MMM-
DD; updated table headers. Changed
<Insert module title> to <Insert
Course code and title>. Deleted
“Healthcare (“Philips”) in legal
statements and made footer
applicable to all pages.

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 7 of 7