VI 500 Weighing Indicator Manual

This document provides instructions for using a VI 500 weighing indicator. It discusses safety instructions, functions of the indicator including weighing, printing, and optional features. It also provides details on installation, calibration, operation, maintenance and troubleshooting. Key steps outlined include connecting the load cells and power supply, entering settings mode, following the calibration procedure, and restoring default parameters when needed.

Uploaded by

Juan David
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
321 views31 pages

VI 500 Weighing Indicator Manual

This document provides instructions for using a VI 500 weighing indicator. It discusses safety instructions, functions of the indicator including weighing, printing, and optional features. It also provides details on installation, calibration, operation, maintenance and troubleshooting. Key steps outlined include connecting the load cells and power supply, entering settings mode, following the calibration procedure, and restoring default parameters when needed.

Uploaded by

Juan David
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd

VI 500

Weighing indicator

USER MANUAL

Edition:12062801A
Safety Instruction

For safety operation pls. follow the safety instruction.

WARNING

Setting. Calibration Inspection and Maintain of the indicator is


prohibited by Non-professional staff.

WARNING

Pls. make sure the weighing display have good ground in using

WARNING

The indicator is the static and sensitive


equipment, cut off the power during electrical
connections, internal components touched by
hand is prohibited, and please take the
measure of anti-static.

LIST
1. Instruction...............................................................................................................1
1.1 Main function................................................................................................1
1.2 technical parameter.......................................................................................1
1.3 drawing..........................................................................................................1
1.4 Battery instruction.........................................................................................2
2.Installation and calibration...................................................................................2
2.1 Power supply connection..............................................................................2
2.2 Connection of load cell and indicator..........................................................3
2.3 Communication interface.............................................................................3
2.4 4~20mA output.............................................................................................4
2.5 Relay output signal function........................................................................5
3. Basic operation.....................................................................................................8
3.1 keypad............................................................................................................8
3.2 Power on & off..............................................................................................9
3.3 Zero operation...............................................................................................9
3.4 Tare operation.............................................................................................10
3.5 Accumulation operation.............................................................................10
3.6 Print..............................................................................................................10
3.7 Hold..............................................................................................................11
3.8 Count............................................................................................................11
4. Calibration and Parameter setting.................................................................11
4.1 Enter setting.................................................................................................11
4.2. Step of calibration operation:....................................................................12
4.3 Application function parameters setting chart..........................................14
5. Output format......................................................................................................19
5.1 Big display continuous sending format.....................................................19
5.2 Computer continuous sending format.......................................................20
5.3 Serial interface reception command..........................................................21
5.4 Print format..................................................................................................21
5.5 PC or Big display continuous sending format..........................................22
6. Maintenance........................................................................................................23
6.1 Regular error and solution..........................................................................23
6.2 Daily maintain.............................................................................................24
6.3 Restore default parameter...........................................................................24
6.4 Packing list..................................................................................................27
1. Instruction
This weighing indicator is designed for bench scale, floor scale. The basic weighing function include: Hold, Print .
Optional: I/O , 4-20mA output.

1.1 Main function

Weighing function:
Zero, tare, accumulation, printing, animal weighing.
Overload remind.
Print format: Date, Time, Net,Tare,Gross
Options:
Pinter
RS232/RS485 serial interface or second display
I/O
4~20mA

1.2 technical parameter


Accuracy class 6000 e

Resolution display: 30, 000 ADC: 2,000,000


Zero stability error TK0 < 0.1μV//K
Span stability error TKspn < ± 6 ppm//K
Sensitivity (internal) 0. 3 μV /d
Input voltage -30~30mV DC
Excitation circuit 5 VDC, 4 wire connection,
Maximum connect 6 load cell of 350Ω
AC power AC100~250V
Operation temperature -10°C~+40°C
Operation humidity ≤90%RH
Storage temperature -40°C~+70°C
1.3 drawing

1
LP7510
Weighing
Indicator
Hi

ok

Gross Net Tared Total Lo

Hold pcs
kg
Battery
ON
OFF Total Gount kg/lb Gross Tare
Zero Print
L O S
C
CO
Class
Capacity= d=

LP7510 Weighing Indicator


Hi

ok

Lo

Gross Net Tared Total Hold pcs kg Battery

ON Total Gount kg/lb Gross Tare Zero Print

OFF

Class Capacity= d=

1.4 Battery instruction

1. when you use the internal battery first time,you should charge the battery fully, to prevent low
voltage resulted from self leakage of battery.
2. when the “red” light is flash, means low battery, pls. charge it in time
3. the light turns to red during charging
4. when the light turns to green, means fully charged.
5. if battery is not used for long time, take it out to avoid the leakage.
6. In order to keep the battery in best using condition, it is suggest that you fully discharge the battery
every month, the method is that using the indicator till it is automatically power off.

2.Installation and calibration

2.1 Power supply connection

The indicator is powered by adapter, you plug the adapter directly into the “DC” pin at the back cover the
indicator is ok.

2
2.2 Connection of load cell and indicator

The indicator can connect with 6 load cell of 350Ω at most, 4 wire or 6 wire load cell both ok.
There are two methods connection between load cell and indicator
A. quick disconnect, as below:

B. Terminal trip connection (inner connection)


1. the exciting voltage for the load cell is 5VDC, the largest output current 120mA, maximum connect 6
pcs 350Ω load cell;
2. Load cell (or the signal cable for the junction box) is connected with 5 bit terminal trip ( SENSOR)
on the printed circuit board of weighing indicator.
3. Open Weighing indicator back cover, insert signal cable to the terminal trip(SENSOR), and make
sure the screw is fixed tightly, the connection as below:

Weighing indicator Load cell


+EXC + voltage

+SIG + output signal


HD ----------- Shielded wire
-SIG - output signal
+EXC +SIG HD -SIG -EXC -EXC - voltage

2.3 Communication interface

RS232 : DB9 Pin or 3 Pin

DB9 definition

5 4 3 2 1
9 8 7 6

Pin function and definition as bellows:


3
DB9 joint Definition Function
2 TXD Sending data
3 RXD Receiving data
5 GND Ground interface
Note: if RS485, The connection pin is 2 and 5 pin.

3 Pin definition

connection:

TXD RXD GND A B


Pin definitions:

Pins Definitions Function

TXD Sending data


RS232
RXD Receiving data

GND Ground

RS485 A RS485 output “A”port

B RS485 output”B”port

2.4 4~20mA output

Technical parameters:
Resolutions: 1/1000
Outside Load: 100-350Ω
4
Connection:
1. Inside connection: 4-20mA, load input port connect with “I” of J602, Ground port connect with “ GND” of J602
2. Outside connection: 4-20mA, load input port connect with “1” pin of DB9, Ground port connect with “6” pin of
DB9 Testing :

Connect the 250Ω to the 4-20mA, adjust the multimeter to the current stalls, the red pan connect the loading port, and the
black pan connect” GND” to test the output current.

Calibration:
1. Press” Print” and “ Total” go to C32, show[out-4], the output current should be 4mA.
2. If Press [↑]show[out-5],Pres[↓]show[out-20],the output current should be 20mA.

3. Adjust the current, for example, adjust to [out-20], Press [←](“Zero” key)or[→](Net/gross key)to adjust the

current Function instruction:

4~20mA correspond to Zero ~Max. capacity. After press” TARE”, the weight will start from 4mA. SET C31=0, 0~20mA
output mode.; C31=1 4~20mA output mode.

If you need 0~5V output, Set C31=0, Then connect 250Ω at the two ends of current output., will get 5voltage at resistance
two ends.

2.5 Relay output signal function

The indicator can output 4 signal , connect with the outside equipment the indicator can perform automatic
control function and upper limit and lower limit alarm function. Perform the 4 kinds function through setting
C33, 4 signals

As below

Output Port definition Function


port
Out1 Close output function No output signal
C33=0 Out2 Close output function No output signal
Out3 Close output function No output signal
Out4 Close output function No output signal
Out1 Open overload control function Output overload
control signal
Out2 Open compliance control function Output compliance
C33=1 control signal
Out3 Open underload control function Output underload
control signal
Out4 Open stable control function Output stable control
signal
5
Out1 Weight>=C13 instant connection Feeding control
signal
Out2 Weight<=C14 instant connection Feeding control
signal
C33=2 Out3 OUT1、OUT2 close When OUT1、OUT2
coneec will output
alarm signal,
connect with buzzer
Out4 Weight<=C14 constant connection Feeding start and
Weight>=C13 disconnect stop
C33=3 Preserved, no function.

For example:

Check weigher application. Connect indicator with yellow, green, red 3 lights. Yellow light on when
overload, if ok the green light on. If underload red light on. And can connect with buzzer. There would be
alarm remind when overload.
C33=2 Feeding control application

For hopper scales. The target is between is 50~500kg, the I/O card can perform open and close feeding
valve.
Parameter setting C13=500kg C14=50kg

C13=500kg upper limit , close feeding valve

C14=50kg lower limit, open feeding valve.

Connection: 1. Connect directly control machine, pls. connect OUT 4, suggest connect another
manually switch as the stop switch in emergency
2. indicator output signal will influence the ON/OFF switch to control machine, OUT 2 ON, OUT 1
OFF, OUT 3 connect buzzer, alarm remind when start and stop.
Instruction: when the weight over 500kg, OUT 1 OU3 instant connection and break off, OUT 4 break
off; when weight under 50kg, OUT 2 OUT3 instant connection and break off, OUT 4 constant
connection.

5 4 3 2 1
9 8 7 6
DB9 pin definition port

1 pin output signal pin Out1


1st
6pin output signal pin Out1
1st
2 pin nd Out2
2 output signal pin
7pin nd Out2
2 output signal pin
3 pin rd Out3
3 output signal pin

6
8 pin rd Out3
3 output signal pin
4 pin th Out4
4 output signal pin
9 pin th Out4
4 output signal pin
Inner connection pin definitions

01 01 02 02 03 03 04 04

7
3. Basic operation

3.1 keypad

LED instruction
Weighing data

Hold Data hold


Gross Gross weight
Net Net weight
Tare tare
The weighing data is stable
Weight is zero
Hi Overload
OK ok
Lo Underload
. Decimal
TOTAL Go to accumulation mode

8
Keys function
keys Key name Key function

Print Print

Zero Zero the weight within tolerance

Tare At G.W mode, get the tare weight. At N.W


mode, clear the tare, get the G.W

Gross weight At N.W mode, check the G.W, after 3 seconds


back to N.W automatically

Holding the 1. Holding the data


2. Work with “PRINT” go to calibration
data
Counting Counting operation

1. Accumulation
Accumulation 2. work together with “ Print” to perform
The accumulation function and check the
accumulation result
Power on/off Press 2 seconds to power on or power off

3.2 Power on & off

Press 2 seconds to power on or power off, after power on the indicator show”000000-999999”. After
self inspection. It go to the weighing mode. Pls. check it whether 6 bits LED/LCD display and the status light is
good or not.

3.3 Zero operation

1. Initial zero setting


When power on the indicator, if the weight on the scale is within the initial zero tolerance, indicator show zero

9
automatically.
2. Manually Zero setting
When the scales is stable, and not the negative display, you can zero the weight within tolerance by press

keys.

3.4 Tare operation

Press “TARE” key, the gross weight is tared, indicator show the Net weight, the “Net” “tared” status light is on.
At tare mode, Press” TARE” key, clear the tare weight, the indicator will show the gross weight.

3.5 Accumulation operation

At Zero mode, load weight till stable, Press go to accumulation

mode, ”Total” light on, display” n 001”, and then display loaded weight; unload the weight , back to zero, load the

second weight again till stable. Press display”n002” then display the second loaded weight. Repeat it agin
and again, maximum 999 times.
Check the accumulation

Press and hold it then press , display ”n**”, (it is the accumulating times) then show total weight.
there are 8 digits totally. It shows the first 4 digits then the last 4 digits. For example, the first 4 digits is”0012”, the
last 4 digits is”34,56” It means the actual weight is “1234.56”
EXIT the accumulation function

When the indicator show the last 4 digits, Press hold it, the indicator show “ clr n”, it means don’t clear
the total Weight, Press “PRINT” key to exit it; if you want to clear total weight, Press “ZERO” or “TARE” key, “clrn”
change to “clry” it means clear total weight ,then Press “PRINT” to clear the total weight and exit accumulating
mode.

3.6 Print

If the weighing is stable, after connect with printer, press” PRINT” can print the weight. Note: at tare
mode, print with tare. if negative weight,, can not print. Set C30 for time format.
10
3.7 Hold

There are two different hold function. Peak hold function and data hold function. And the setting is different
accordingly.
option:0=close hold function
1=Peak hold /2=Data hold
3=Auto hold/4=Animal hold
Instruction:
Peak-hold: it shows the max. data,
mainly application for materials testing, such as tension and pulling force.
Date-hold: it shows current weight value. Mainly application for animal weighing.
Auto-hold: When the pet keep stable on the scale, the indicator will automatically “ Hold” for 6 seconds.
Animal-hold: Press” Hold” key, the indicator will show” LOC” for 3 seconds, the “ hold” light is on,
During the 3 seconds, the indicator will catch the average weight and show it.
Press” HOLD” key again to exit it

3.8 Count

1. At weighing mode, load the weights on the platform scales, Press” Count” the indicator show” PCS 0”
press” Zero” key input the quantity, press” Print” to confirm it
2 . Load the goods on the platform scales, then the indicator will show the quantity.
3. Press” Count” back to weighing mode.
4. If you want to weigh different goods, at weighing mode, put the sample on the platform scales, press”
Count” the indicator show”0”
Press “Print” hold it and then press “Count” the indicator show” PCS 0”, press “Zero” input the sample
quantity, press “ Print” to confirm it. Then repeat the step 2 and 3.

4. Calibration and Parameter setting

4.1 Enter setting

There have two methods to enter the setting menu:


1. Put the switch “MARK” to “On” position, enter calibration
And when calibration is finish. Put the switch “ MARK” to “off” position.
Then add the sealing screw at the back of indicator.

2. when the switch “ MARK” is on position, press the “ ” hold it and then press” ” enter

calibration.

11
The key functions in setting:

ENTER

UP
DOWN
LEFT
RIGHT
BACK
EXIT CALIBRATION
POWER

4.2. Step of calibration operation:

According to the second method which can enter setting menu, C01-C39

step Method of operation display Remark


1 [C01 ] After you enter calibration
mode, it display [C01 ]
2 press [C01 1] Weight unit kg

3 press [C02 ] Set decimal digits


press [C02 0] option:0/1/2/3/4
press or [C02 2] Select decimal digit
example:two decimal point:
[C02 2]
4 press [C03 ] Set graduation
press [C03 1] option:1/2/5/10/20/50
press or [C03 5] Select required graduation
example:graduation 5:[C03 5]
5 press [C04 ] Max capacity
press [0100.00]
press or / [0100.00] example:max weighing 100kg:
12
[0100.00]
press [C05 ] Zero calibration
press [C05 0] Option
press [C05 1] 0=no need zero calibration
press [CAL 9] 1=need zero calibration
。。。。。。 calibration zero please choose 1
6 [0000.00] and ensure scale is empty and
“stable” light is on
Ensure zero calibration,
countdown. Till
show[0.00](example for two
decimal point)。
7 press [C06 ] calibration
press [C06 0] option:
0=No need calibration
press or [C06 1] 1= need calibration
Load weights on scales
press [SPAN ] according to max. capacity.
[0100.00] Suggest close to the max
capacity, at least 10% of max.
capacity.
press or [0080.00] For example: the weights is
press [CAL 9] 80kg
…… As bellows:
[0080.00] Input the 0080.00, count down ,
[CAL End] then indicator shows 0080.00 ,
calibration is over.
If you want to set application
function parameter. Press
“PRINT” if you want to exit
press “HOLD”
press [C07 ] Default parameters setting
8 press [07 0] option:0=non-restore default
press or [07 1] parameters
1=restore default parameters
Note: after the above
parameters setting finish, please
do not set default parameters to
avoid the original setting
parameters is lost.
13
4.3 Application function parameters setting chart

Function Setting parameters setting and instruction


Item
warning C08 Options: 0 = close warning tone
warning 1 = open warning tone
tone
tone
option:0=close auto power off
10= power off automatically if no
Automatic C09 change within 10 minute.
Automatic 30= power off automatically if no
power off
power off change within 30 minute.
60= power off automatically if no
change within 60 minute.
LED Version:
option:0= close power saving setting
C10 3= close display if no change within 3min.
Power 5= close display if no change within 5 min.
Power
saving LCD Version:
saving
setting 0=Close he backlight
setting
1= backlight when the weight change or
press the keyboard
2=constant backlight
option:0=close hold function
1=Peak hold /2=Data hold
3=Auto hold /4=Animal hold
Hold C11 Instruction:
Hold mode Peak-hold: it shows the max. data,
function
mainly application for materials testing,
such as tension and pulling force.
Date-hold: it shows current weight value.
Mainly application for animal weighing.
C12 option:3=3 second
Animal hold Animal hold 5=5 second
sample time sample time

C13
Upper/lower Upper limit You can set it within the max. capacity limit
limit alarm alarm value

14
C14
Lower limit
alarm value
Inner Code C15
display Check inner enter C15 to check the inner code
code

C16 Enter C16, you can set the date,


Date and Date from left to right: year/month/day
time C17 Enter C17, you can set the time from
Time left to right: hour/min./sec.
option:0= Close serial interface data
output
1=Continuous sending, connect big
display
C18 2=Print method, connect printer.
Communica Serial interface data 3=Command request method ,
output method connect computer.
tion setting
4=PC continues sending format,
connect computer.
5=PC/ big display continuous
sending format.
C19 option:
Baud rate 0=1200/1=2400/2=4800/3=9600
Option:
C20 0= close manually zero setting
1=±1% max capacity
Manually zero range
2=±2% max capacity
Zero range
option:0= no initial zero setting
C21 1=±1% max capacity
2=±2% max capacity
Initial zero range
5=±5% max capacity
10=±10% max capacity

15
Options:
0= close zero tracking
0.5=±0.5d
1.0=±1.0d
C22 2.0=±2.0d
Automatically zero 3.0=±3.0d
tracking range 4.0=±4.0d
5.0=±5.0d
Zero Note: 1. d = division

tracking 2. the zero tracking range can


not bigger than manual zero range.
C23 Options:
0= close zero tracking time
Automatically zero 1= 1 second
tracking time 2= 2 seconds
3= 3 seconds

Overload C24 option:00= close overload range

range Overload range 01d~99d


remark:d =division

Negative C25 Option:0=-20d


display Negative display range 10=10% max. capacity
20=20% max. capacity
50=50% max. capacity
100=100% max. capacity
Standstill C26 Option:
time Standstill time 0= quick 1= medium 2= slow

C27 Option:
Standstill range 1= 1d 2=2d 5=5d 10=10d
D= division

Digital filter C28 option:0= close dynamic filter


Dynamic filter 1=1 digital filter strength
Instruction : Dynamic 2=2 digital filter strength
filter is collecting the 3=3 digital filter strength
data filter before loaded 4=4 digital filter strength
weight stable. When 5=5 digital filter strength
loaded weight easily 6=6 digital filter strength
16
shaking (for example Note : Pls setting dynamic filter
animal) , you can set strength carefully, the No. is
this filter to make bigger, more stable. if the
weight display more loaded weight shake not too
stable much. The setting is less than 3

C29 option:0=close noise filter


Noise filter 1=1 digital filter strength
2=2 digital filter strength
3=3 digital filter strength

C30 C30=0 yy.mm.dd


Print time and date C30=1 mm.dd.yy
C30=2 dd.mm.yy
C30=3 yy.mm.dd

Analog C31 output type C31=0 0~20mA ouput


output C31=1 4~20mA output
setting
4~20mA C32 calibrate current Refer to 2.5
current
calibrate
Relay C33 Relay output C33=0 close relay output
output C33=1 Open relay output function 1
setting C3=2 Open relay output function2
C33=3 Preserved menu
Muti C34 Communication C34= 0~99 Add. Code
communicat add.
ion add.
Wireless C35 C35=0~99 signal
communicat
ion
Gravity of C36 C36=9.7000~9.9999
calibration
location
Gravity of C37 C37=9.7000~9.9999
destination
Version No. C38
Preserved C39
menu
17
Print mode C41 C41=0 auto mode
C41=1 gross mode
C41=2 tare mode
See 5.4 Print format parts in detail
Print C42 C42=0~9
carriage
return
Space Print C43 C43=0~9

Date Print C44 C44=0 not print date


C44=1 print date
Time Print C45 C45=0 not print time
C45=1 print time

18
5. Output format

5.1 Big display continuous sending format

Output continuous format


S S S S C C
T W W W X X X X X X X X X X X X K
R
X A B C S

1 2 3 4 5 6

State A
Bits0,1,2
0 1 2 Decimal point position
0 1 0 XXXXXXX
1 1 0 XXXXX.X
0 0 1 XXXX.XX
1 0 1 XXX.XXX
Bits3,4 Division
0 1 X1
1 0 X2

State B
BitsS function
Bits0 gross=0, net=1
Bits1 Symbol: positive =0,negative =1
Bits2 Overload(or under zero)=1
Bits3 dynamic=1
19
Bits4 unit:lb=0, kg=1

Bits5 Constant 1
Bits6 Constant 0

State C
Bit2 Bit1 Bit0 unit
0 0 0 Kg or lb
0 0 1 g
0 1 0 t
Bit 3 printing=1
Bit 4 Extend
display=1

Bit 5 Constant 1
Bit 6 Constant 0

5.2 Computer continuous sending format

, , CR LF
S1 S2 S3 Data S4

S1: weight status, ST= standstill, US= not standstill, OL= overload

S2: weight mode, GS=gross mode, NT=net mode


S3: weight of positive and negative, “+” or ” –“
S4: “kg” or “lb”
Data: weight value, including decimal point
CR: carriage return
20
LF: line feed

5.3 Serial interface reception command

RS232COM serial interface can receive simple ASCII command.


Command word and role as follows:
Command NAME Function
T TARE Save and clear tare
Z ZERO Zero gross weight
P PRINT Print the weight
R Reply Reply the command
C Kg/lb Kg/lb conversion
G G.W Check gross weight at net weight mode

R command receive data format

5.4 Print format

Tare mode:
21
Date: XX.XX. XX
Time: XX:XX:XX
NET XX.X kg
TARE XX.X kg
GROSS XXX.X kg
Gross mode:
Date: XX.XX. XX
Time: XX:XX:XX
GROSS XXX.X kg

5.5 PC or Big display continuous sending format

22
6. Maintenance

6.1 Regular error and solution

ERROR REASON SOLUTION


1. Overload 1. reduce the weight
2. wrong connection with 2. check load cell connection
UUUUUU load cell 3. inspection load cell. Check the
3. load cell has quality input and output
problem.

1. wrong connection 1. check load cell connection.


nnnnnnn 2. load cell has quality 2.check input and output resistance
problem to judge it is good or not.

During calibration, not Input the correct weights


ERR1 input the weights or the
weight is overload
During calibration , the The calibration weights Minimum is
ERR2 weights is below than 10% of Max. cap. Recommend
Min. required weights 60%-80% of Max. Cap.

1. check the connection is correct


ERR3 During calibration, the 2. check load cell is no problem
input signal is negative 3. recalibration if still wrong
change the PCB

23
ERR4 During calibration, the After the platform is stable, start
signal is unstable calibration
ERR5 EEPROM error Change PCB

ERR6 Exceed zero range Remove the load

6.2 Daily maintain

1. Protect the indicator from strong sunlight to prolong the using life

2. Good connection between load cell and indicator. Far from away
from strong electric field, magnetic field.
3. Power off the indicator when lightning
4. Power off the indicator firstly before plug and unplug

6.3 Restore default parameter

Enter to calibration, Set C07=1. Press” ” then press” ” to exit


saving setting. All parameter will back to default

Note : Pls. do not restore default parameter easily if you are not
professional staff or not yet calibrate the scales.
Default parameter

Parameter instruction Default


C01 Calibration 1
C02 Decimal digits 0
24
C03 Resolution 1
C04 Max. capacity 10000
C05 Empty calibration 0
C06 Capacity calibration 0
C07 Restore default 0
C08 Warning tone 1
C09 Power-off automatically 0
C10 Power saving mode 0
C11 Hold function 0
C12 Animal hold sample time 3
C13 Upper limit alarm 000000
C14 Under limit alarm 000000
C15 Inner code
C16 Date setting
C17 Time setting
C18 Serial interface data output 0
C19 Serial interface Baud rate 3(9600)
C20 Zero manually 10
C21 Initial zero 10
C22 Zero tracking range 0.5
C23 Zero tracking time 1
C24 Overload range 9
C25 Negative range 10
C26 Standstill time 1
C27 Standstill range 2
C28 Dynamic filter 0
C29 Noisy filter 2
25
C30 Print format 0
C31 Analog signal options 1
C32 4~20mA testing 4
C33 Relay output setting 1
C34 Muti PC communication add. 0
C35 Wireless communication channel 6
C36 Calibration location gravity 9.7936
C37 Destination gravity 9.7936
C38 Version No. check
C39 Reserved menu

26
6.4 Packing list

Packing list

S/N ITEM NAME UNIT QTY PACKING

1 Weighing PCS 1
indicator

2 Plastic bag PCS 1


3 bag PCS 1
China/DC9V PCS 1
US/DC9V PCS 1
4 Adapter UK/DC/9V PCS 1
EU/DC9V PCS 1
AU/DC9V PCS 1
OTHERS PCS 1
5 USER MANUAL PCS 1
6 RS232 3 PIN OR DB9 PCS 1
7 LOADCELL 5 PIN Quick PCS 1
PLUG disconnect

8 Signal cable Φ5/3 core shield PCS 1


signal cable

9 Power cable 3 coreΦ0.75mm PCS 1


10 Bracket Wall-mounted PCS 1
11 Certificate PCS 1
12 Packing list PCS 1