Product Specifications
1-1. Product Specifications
Frequency
► Range : 9 kHz to 3.0 GHz
► Resolution : minimum 1 Hz
► Span Range : 100 Hz / div to 300 MHz / div
Selection of 1,2,5 steps (automatic), ZERO Span, FULL Span (9 kHz to 3.0 GHz)
► Frequency Selection : Start, Stop, Center, Span set-up
► Span Accuracy : ±3 % of the Indicated Span Width
► Readout Accuracy : ≤±(Indicated frequency × Reference frequency accuracy + Span
× Span accuracy + 50% of RBW)
► Phase Noise : ≤−90 dBc/Hz @10 kHz offset
Amplitude
► Range : +20 dBm ~ −105 dBm
► Avg. Noise Level (1 kHz RBW, 10 Hz VBW)
≤−105 dBm : 150 kHz ~ 1 GHz
≤−100 dBm : 1 GHz ~ 2.4 GHz, 50 kHz ~ 150 kHz
≤−95 dBm : 2.4 GHz ~ 3 GHz
► Amplitude Unit : dBm, dBmV, dBuV, V, mV, uV, W, mW, uW
► Display Scale Linearity
≤±1.5 dB / 70 dB (10 dB / div), ≤±1.5 dB / 40 dB (5 dB / div)
≤±0.5 dB / 8 dB (1 dB / div), ≤±0.5 dB / 16 dB (2 dB / div)
► Frequency Response (0 dB attenuation) : −3.5 dB ~ 1.5 dB (100 kHz ~ 10 MHz)
±1.5 dB (10 MHz ~ 3 GHz)
► Reference Level
Range : −90 dBm to +20 dBm
Resolution : 0.1 dB step
Accuracy : ±1.5 dB
► Second Harmonic Distortion : ≤−60 dBc, −40 dBm input
► Intermodulation Distortion : ≤−70 dBc, −40 dBm input
3
► Residual Spurious : ≤−85 dBm (Input terminated, 0 dB attenuation)
► Other Input Spurious : ≤−60 dBc, −30 dBm input
► Resolution Bandwidth
Selections : 1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz, 300 kHz, 1 MHz, 3 MHz, 9 kHz, 120 kHz
Accuracy : ±20 %
Selectivity : 60 dB / 3 dB ratio < 15 : 1
60 dB / 6 dB ratio < 12 : 1 (9 kHz, 120 kHz)
Switching Error : ≤±1.0 dB (1 kHz Reference RBW)
Video Bandwidth : 10 Hz to 3 MHz in 1-3-10 step
Sweep
► Time : 100 ms to 1000 sec, 40 ms to 1000 sec (zero span)
► Accuracy : ≤±20 %
► Trigger Source : External (rear), Video, Free run, Line
► Trigger Mode : continuous, single
► Trigger Level : TTL level
Storage
► Trace Storage : maximum 900 waveforms
► Setup Storage : maximum 3,000 states
Screen Display
► Type : 6.4″ color TFT LCD screen
► Display Resolution : 640 (H) × 480 (V) active display area
► Marker Mode : Peak search, Delta marker, Marker to Center,
Marker to Reference (8 markers maximum)
Input
► RF Input Connector : N-type Female, 50 Ω nominal
► VSWR : 150 kHz to 3.0 GHz, VSWR < 1.5 : 1 (with 0 dBm Ref Level)
► Maximum Input Level : 0 Vdc, +20 dBm
4
Standard Frequency (10 MHz, Ref.)
► Temperature Stability : ±0.5 ppm
► Aging : ±0.5 ppm / Year
► Connector : BNC female
► Input Level : −5 dBm to +15 dBm
► Output Level : 10 MHz, +8 dBm nominal
Interface
► RS-232C
► Printer
Driver : PCL3 or PCL5, HP, EPSON, Laser-Jet, Desk-Jet etc. supports most of printers
Connector : for standard 25 pin female D-Sub parallel printer, support USB
► USB 2.0 Host
Printer Driver : Supports most of printers including HP, EPSON, Laser-Jet, Desk-Jet, etc.
USB Storage Device : supports 1.1, 2.0, for storing image files, supports GIF format
► Ethernet (Option)
10-Base-T Ethernet : supports Internet remote control
► GPIB Interface (Option) : IEEE 488 bus
General Specifications
► Size : 350 (width) × 195 (height) × 375 (length) mm
► Weight : 10 kg
► Warming-up Time : More than 20 minutes for precise measurement
► Power
Supply Electrical Power : 100-240 VAC at 50 / 60Hz
Consumption Power : 80 watts maximum (when an option is not built in)
► Operating Temperature : 0 ℃ to 40 ℃
► Temperature for Storage : −20 ℃ to 70 ℃
► RF Emissions, Immunity
RF emissions : EN 550011
5
Chapter 2. Preparation for Use
2-1. Initial Inspection
Please inspect all packing boxes and check if they contain all contents. Keep the packing
box and packing materials until the inspection by an analyzer is completed.
There is information on accessories offered along with an analyzer in Table 2-1. Please
contact LP customer support center when the contents are lacking or there is something
wrong with the analyzer.
9
2-2. Requirements for Power
This analyze is a portable device, and does not need the additionally established devices
except for the connection with power.
You need not select line power.
When there is no spare fuse offered along with the device, the fuse to change should
match the fuse in the fuse holder, power, and rating electric current (250 VAC, 3.15 A type
T 5 × 20 mm).
A fuse should be established in the fuse holder located in the upper part of a power switch
on the back board.
To prevent the danger of fire, please use only the same type and rating of
WARNING !
power fuse when changing it. The use of a fuse with different rating may
incur the damage to the device.
10
2-4. Power Cable
The analyzer uses a power cable with three lines in accordance with the international
safety standard. When it is connected to a power socket concerned, this cable grounds
a cabinet of the device.
When there is no grounded AC power socket, FG terminal on the rear panel should be
grounded.
Use a grounded power cable with three lines, or connect the analyzer to
WARNING !
a protective ground.
If you turn on the power without following these warnings, you might have
the danger of electric shock.
Also, you need to check the electric pressure of power. When power
exceeding the standard value, comes into the analyzer, the analyzer might
be damaged or be caught in a fire.
If you use the analyzer in the normal temperature, after using or keeping
WARNING !
it in a low temperature for a long time, there is a danger that it might get
short-circuited due to condensation. Thus, in order to prevent such a danger,
please do not supply power until the analyzer is fully dried.
In order to prevent the inner temperature of the device from rising, there is
WARNING !
a freezing fan on the rear panel of the analyzer. You should leave at least
10 cm between the back panel and a wall or nearby device, or barrier,
so that a ventilating hole does not get clogged.
11
2-6. Turning on Power
Please connect a power code to the analyzer, before using the analyzer.
Please press the On key.
In order to use the analyzer, please check the operating temperature, and heat the analyzer
for about 10 minutes before measuring.
When you intend to use an external 10 MHz signal as a standard frequency, please
connect the external standard signal to a 10 MHz REF IN connector on the rear panel.
A signal level should be over -5 dBm.
12
LPT-3000
Chapter 3. Starting
3-1. Function of a Front Panel
1. It is 6.4″ TFT-LCD.
It indicates wave type, parameter set-up, indicated value, and menu key label.
2. A menu key is beside a screen, and there is no label indicated on it. The menu key label appears in
a form of a button on an LCD screen beside the key. Most of keys of which their labels are indicated
on the front panel of the analyzer, can be used via the menu key with functions concerned.
3. FREQUENCY, SPAN, and AMPLITUDE keys activates the main functions of the analyzer, and make
it possible to use menus with functions concerned.
FREQUENCY key is an input part of frequency parameter data.
SPAN key is an input part of span parameter data.
AMPLITUDE key is an input part of amplitude parameter data.
4. MARKER function controls marker, reading frequency, and amplitude on the trace of the analyzer,
and automatically finds out a signal of the highest amplitude.
MARKER key is for setting a marker.
PEAK SEARCH key is related to the peak search function.
15
MARKER → This key is for setting a marker value as a specific parameter.
5. CONTROL function controls the broadband of resolution and sweep time, and make it possible to
use a menu controlling a device display. Also, it makes it possible to set the variable of other
analyzer necessary for measurement.
I/O DETECT key is for setting input and output of a Ref. 10 MHz signal, and Detect Mode.
TRACE key is related to the set-up of trace on an LCD screen.
AUTOSET key is for searching for voluntary signal, and automatically setting a measuring
parameter.
BW/AVG key is for setting functions including RBW, VBW, and AVG.
DISPLAY key is for setting a screen indicating functions.
TRIG key is for setting a trigger function.
SINGLE key is for performing the single sweep.
SWEEP key is for setting a sweep time and sweep mode.
6. MEASURE makes it possible to use key menus for performing the measurement by the analyzer.
Once the measurement gets started, you can use an additional menu key that defines the set-up of the
current measurement function by using a MEAS SETUP key. MEAS CONTROL and RESTART make
it possible to use additional measurement controlling function.
MEASURE key is for performing the measurement function.
MEAS CONTROL is for controlling the measurement function.
MEAS SETUP key is for setting the measurement function.
RESTART key is for restarting the measurement.
7. The function within a SYSTEM key, affects the entire state of the analyzer. You can make diverse
set-up and arrangement by using this SYSTEM key.
PRESET key is initializing the analyzer to an established state.
FILE key menu makes it possible to save the current trace state in a memory of the analyzer or USB
flash memory, or make it to be loaded from them.
SAVE key immediately perform the currently defined saving function.
PRINT SETUP menu key makes it possible to set printing stuffs.
PRINT key immediately sends printing data to a printer.
8. This is a power switch. It is used when the main power switch on the back panel is on. If you lightly
press the key, there comes the power under STBY condition. If you click the key for three seconds,
16
it can revert to a STBY condition from the state of turning on of the power.
9. This is a USB Connector for USB Flash memory and USB printer.
12. A data controlling key including a step key, knob, and number key pad, makes it possible to change
the value of the currently activated functions.
The data controlling key is used for changing the value of functions including central frequency,
starting frequency, resolution band, and a marker location.
Data controlling changes the current functions in a stipulated way of functions concerned. For
instance, the central frequency may adjust value precisely by using a knob, or change the value to
be accurate value by using a step key or number key pad step by step.
The knob makes it possible to successively change the functions such as central frequency,
reference level, and a marker location. When you turn the knob clockwise, the value will go up. As
for making successive changes, the scope of adjustment will be determined upon the scale of the
scope of measurement.
The knob makes it possible to change central frequency, starting or suspending frequency, or
reference level. When the central frequency or reference level is adjusted, a signal may move from
the right to the left, or goes up or down in accordance with the revolution of the knob before a new
sweep is actually performed.
The number key pad is used for inputting accurate value on the currently established functions of
the analyzer. A decimal point might be used with a number part. If not, the decimal number will be
located at the end of the number.
The input of a number will be completed only when pressing a unit key. Once the input of the
number gets started, a unit key label will appear on a menu key. The unit key will be changed upon
the currently activated function. For instance, unit keys on frequency span are GHz, MHz, kHz, and
Hz, and a unit key of reference level is dBm.
17
3-2. Function of a Rear Panel
4. RS-232C connector
6. Input/Output connector of the standard frequency, When the external standard frequency is inputted
in this connector, a user may be able to use this port by operating keys on a front panel.
8. GPIB connector
18
3-3. Explanation of a Display
The below is an example of an item that might be shown on a display of the analyzer.
Please refer to the below explanation for the explanation for a display item. A function key
indicates a key setting a function conforming to the item concerned.
19
Item Explanation Function Key
10 Menu Subject to changes in accordance with the selection of a key
Frequency span or Stop SPAN or
11
frequency FREQUENCY >> Stop Freq
12 Sweep time SWEEP >> Sweep Time
13 Video bandwidth BW/AVG >> Video BW
14 Frequency offset FREQUENCY >> Freq Offset
15 Resolution bandwidth BW/AVG >> Res BW
Center frequency or starting FREQUENCY >> Center Freq or
16
frequency FREQUENCY >> Start Freq
17 Trigger / Sweep TRIG, SWEEP
18 Trace mode TRACE
19 Video avg. BW/AVG >> Average
20 Display line DISPLAY >> Display Line
21 Amplitude offset AMPLITUDE >> Ref Level Offset
22 Amplitude scale AMPLITUDE >> Scale/Div
20
3-4. Outline of a Key
All key labeled as FREQUENCY, SYSTEM and MARKER, are an example of a key on
a front panel.
If you press a key on the front panel, a menu will appear on the right of the display LCD.
A menu shows functions other than functions that can be directly used by a key on the
front board. If you intend to activate a menu concerned, please press a menu key concerned.
The menu indicated will be changed in accordance with a key selected on the front panel.
When the value of the function concerned of the menu can be changed, it can be said that the
function concerned becomes activated. The label of the activated function will be emphasized
after the menu key concerned is pressed. For instance, please press AMPLITUDE key. Then, the
function labeled with Ref level, which is a basic selection key on the amplitude menu will be
activated. Next, Ref Level will show up on an activated function block, and you can input data.
Menus with the indication of on and off can be used for turning on and off of the
functions concerned of the menus. If you intend to activate the function, please press a menu
key so that there appears an underline below On. If you intend to inactivate the function,
please press a menu key so that there appears an underline below Off.
Functions with the indication of Auto and Man can manually change value or can be
automatically connected with other functions. The value of functions can be manually
changed by using a number key pad, scroll knob, or step key. If you intend to automatically
connect the functions, please press a menu key so that there can be an underline below Auto.
21
3-5. Measurement Methods
The 80 MHz standard signal of the inside of an analyzer can be used as a test signal.
1. First, please turn on the device by pressing On. Please wait until the work for power approval and
alignment are completed.
2. Please press SYSTEM >> Power On/Preset >> Preset Type >> Factory in order, and select
a Factory Preset.
3. Please press a PRESET key. Turn on the 80 MHz standard signal of the inside of the analyzer by
pressing I/O DETECT >> Ref Out (On).
22
figure 3-4 fully shows the relationship between the central frequency and reference level. The box
in the figure shows a display of the analyzer. The change of the center frequency changes the
horizontal location of the display. The change of reference level changes the vertical location of a
signal on the display. If you increase a span, the scope of frequency shown horizontal on the display
will increase.
23
3-6. Summary of Measurement
1. Please press SYSTEM >> Power On/Preset >> Preset Type >> Factory in order. Please press a
PRESET key. Turn on the 80 MHz standard signal of the inside of the analyzer by pressing I/O
DETECT >> Ref Out (On).
2. Please set the center frequency by pressing the following key : FREQUENCY, 8, 0, MHz
4. The 80 MHz standard signal of the inside of the analyzer is about -40 dBm. Please adjust reference
level to -30 dBm. If necessary, please activate reference level by pressing an amplitude key, and
locate the signal on the upper side of a screen by changing reference level by using a scroll knob or
step key.
5. Please determine the amplitude and frequency of a signal. You can move a marker to the maximum
value of a signal by pressing PEAK SEARCH or MARKER. Please read amplitude and frequency.
It should be indicated as the figure 3-5. The frequency is indicated horizontally, and the amplitude
vertically.
24
Chapter 4. Menu Structure
4-1. Frequency
4-2. Span
27
4-3. Amplitude
4-4. Marker
28
4-5. Peak Search
4-6. Marker →
29
4-7. I/O DETECT
4-8. Trace
30
4-9. BW/Avg
4-10. Trig
31
4-11. Display
32
4-12. Sweep
4-13. Measure
33
4-14. Measure Control
34
4-16. System
35
4-17. File
36
4-18. File ( Save )
37
4-19. File ( Load )
38
4-21. File ( Copy )
39
4-23. File ( Create Dir )
40
Chapter 5. Explanation of a Key on the Front Panel
5-1. AMPLITUDE
It makes it possible to activate the function of changing reference level, and change
menus related to amplitude. The amplitude menu indicates or corrects data on a vertical axis.
■ Ref Level
You can change reference level. This function is basically activated when you press an
amplitude key. Reference level is a value expressed on the upper part of a lattice line on
a display. If you change the value of reference level, the reference level indicated on the
upper part of a lattice line on the display will be changed. You can change reference level
by using a step key, a scroll knob, or a number key pad. If you press any number among
0 ~ 9 on a number key pad, a unit menu of the current reference level will appear.
AMPLITUDE >> Ref Level
■ Scale/Div
It sets the log unit on the vertical lattice sector on a display. The scope of Scale/Div value is 1 dB ~
20 dB per vertical lattice sector, and can be changed by 1 dB unit. You can change the input
attenuation value by using a step key, a scroll knob, or a number key pad. If you press
any number among 0 ~ 9 on a number key pad, a unit menu will appear.
AMPLITUDE >> Scale/Div
43
■ Ref Level Offset
It adds the offset value to the currently indicated reference level. The reference level
offset value can be inputted only on the same number key pad. If you input the reference
level offset value, trace or input attenuation value will not be changed. The reference
level offset is used when there is a gain or loss of level between the inputs of a device
under test, and the analyzer. Accordingly, a signal level measured by the analyzer can be
thought to be the input level of the device for changing external amplitude. When the
reference level offset is inputted, the value will show up below the offset on the left of the screen. (it is
different from the frequency offset indicated below the lower part of a display.) In order to remove
offset, please press Ref Level Offset, 0, dB. Also, if you press Preset, the reference level offset value will
be established as 0 dB.
AMPLITUDE >> Ref Level Offset
■ Y Axis Units
It displays a menu for changing an amplitude unit. You can change the amplitude unit by
pressing dBm, dBmV, dBμV, Volts or Watts.
AMPLITUDE >> Y Axis Units
44
5-2. BW/Avg
It activates the resolution bandwidth function, and sets the bandwidth function and trace
average function.
45
change the ratio by using a step key, a scroll knob, or a number key pad. If you press any
number among 0 ~ 9 on a number key pad, a unit menu will appear.
BW/AVG >> VBW/RBW
■ Average On Off
It turns on or off the trace average function. When the trace average function is on, and
the detector is in automatic mode, it will be changed to the sample mode.
BW/AVG >> Average
■ EMI Res BW
It makes it possible to select among the resolution bandwidth of 120 kHz, 9 kHz. This
function is established to be none, when the resolution bandwidth is established with
other value by using a Resolution BW key.
BW/AVG >> EMI Res BW
▪ Video
After taking the average of data in the current trace memory, put it on the trace memory
again, and display it. When the average type video is selected, VAvg will be indicated on the
left of display LCD.
BW/AVG >> Avg Type >> Video
▪ Pwr
After taking the average by converting data in the current trace memory, to a linear power
level, then convert it again to a log scale, and then display it. When the average type is
selected, PAvg will be indicated on the left of display LCD.
BW/AVG >> Avg Type >> Pwr
46
5-3. Display
It is a menu setting the parts indicated including a title, display line, and the indication of
lattice, on the analyzer.
■ Full Screen
It makes the measuring window indicated all over the display. If you press a key
indicating a new menu, the full screen function will be cancelled.
DISPLAY >> Full Screen
■ Active Position
It makes it possible to change the location of the activated function block on a display.
DISPLAY >> Active Position
■ Title
It makes it possible to change or delete the title of a display.
▪ Change Title
It enables you to write a title on a title indication line on the upper part of a display. If you
press Change Title, the Alpha Editor menu including usable letters and signs will appear. The
title of a display will be kept until you press the change title again, or withdraw trace saved
along with the previous title.
You can delete the display title by using the delete function.
DISPLAY >> Title >> Change Title
47
▪ Clear Title
Delete the title on the title indication line on the upper part of a display. Once deleted,
it cannot be restored again.
DISPLAY >> Title >> Clear Title
▪ Graticule On Off
It turns on or off the display.
DISPLAY >> Graticule
5-4. File
It is a menu usable for loading, saving, and managing data from the memory inside the
analyzer. The file menu key opens a conversation box equivalent to the function selected.
■ Save
It enables you to save the analyzer state, trace, and screen data on the internal memory.
If you intend to save the file, please perform the following steps.
1. Pleases select data type to save by pressing a menu key concerned. As for trace, you should
select 1. 2. 3 or all on a source menu key.
3. If you intend to change a file name to save, please change the name by using a name menu key.
4. Please press a save now menu key after determining a location to save. When changing a
directory to save, please change the location to save by using a step key and a dir select menu
key.
▪ State
The state file saves the set-up of a device. You can save the analyzer state on a memory with
State.
▪ Trace
Trace can be saved independently or along with the analyzer state. Although CSV type can
48
be read in a form of a spread sheet of a PC, it cannot be drawn to the analyzer. If you save
trace 1. 2 or 3, the trace selected can be saved along with the state. If you save the trace as
All, the trace are all saved along with State in a .trc single file.
▪ Screen
It enables you to save a screen image on a file, and to select GIF and Reverse GIF file format.
The screen image cannot be read with the analyzer.
■ Load
You can load the analyzer state and trace from the internal memory to the analyzer.
If you intend to load a file, please perform the following steps.
1. Put a select bar on a file to load by using a step key. You cannot load GIF file by the analyzer.
As for trace file, please select trace 1. 2 or 3 on Destination.
2. If you are ready to load, please press a Load Now menu key.
▪ State
The state file includes the set-up of the analyzer. If you load the state, the most of set-up is
restored to the value previously saved.
▪ Trace
Trace can be loaded along with the analyzer state when saved. If you load Trace, it will be in
View mode. If you load Trace, it will go to Trace 1 except for Trace All returning to a
location of the state that each trace is saved. Trace + State type only can be loaded.
■ Delete
It deletes a file as a memory of the analyzer.
■ Copy
It copies a file on other directories of the internal analyzer or USB flash memory.
■ Rename
It changes a file name.
49
■ Create Dir
It creates a new directory on the internal memory of the analyzer.
5-5. FREQUENCY
It activates the center frequency set-up function, and displays the menu of the frequency
function. The value of the center frequency or starting and stop frequency are shown below
the lattice of the display.
The analyzer allows the frequency input of the scope exceeding the designated frequency.
However, it prohibits the use of frequency exceeding the scope of frequency of the analyzer.
■ Center Freq
It sets the center frequency. The center frequency can be changed by using a step key,
scroll knob, or a number key pad. . If you press any number among 0 ~ 9 on a number
key pad, a unit menu will appear.
FREQUENCY >> Center Freq
■ Start Freq
It sets the start frequency. The left and right of the lattice conform to the start and stop
frequencies. When the function of changing this start frequency gets activated, the start
frequency and stop frequency will be indicated on a location indicating the center
frequency under the lattice and a span. The start frequency can be changed by using a
step key, scroll knob, or a number key pad. If you press any number among 0 ~ 9 on a
number key pad, a unit menu will appear.
FREQUENCY >> Start Freq
■ Stop Freq
It sets the stop frequency. The left and right of the lattice conform to the start and stop
frequencies. When the function of changing this stop frequency gets activated, the start
frequency and stop frequency will be indicated on a location indicating the center
frequency under the lattice and a span. The stop frequency can be changed by using a
step key, scroll knob, or a number key pad. If you press any number among 0 ~ 9 on a
number key pad, a unit menu will appear.
FREQUENCY >> Stop Freq
50
■ CF Step Auto Man
It changes the center frequency step. The scale of the center frequency step is changed,
and the center frequency button on the frequency menu is emphasized, thus, the center
frequency function gets activated. Then, a step key will change the center frequency in
accordance with the scale of the center frequency step. The center frequency step
function is useful for finding out harmonics or side band exceeding the current frequency
span of the analyzer. When selecting Auto, the scale of the center frequency step will be established
as one lattice (10% of a span). It can be changed by using a step key, scroll knob, or a
number key pad. If you press any number among 0 ~ 9 on a number key pad, a unit
menu will appear.
FREQUENCY >> CF Step
■ Freq Offset
It enables you to input the frequency offset value. The offset can be inputted only in a
number key pad. Even if you input the offset, there is no change in trace. If you intend to
remove the offset, please press Freq Offset, 0, Hz. When the frequency offset is inputted,
the value will show up on the lower part of the display. (it is different from the reference
level offset indicated on the left of the display.)
FREQUENCY >> Freq Offset
51
5-6. I/O DETECT
■ 10 MHz
It selects the use of the 10 MHz standard signal of the analyzer. When Int(Internal) is
selected, use the standard signal of the inside of the analyzer, and output the 10 MHz
standard signal with a port on the back. When Ext(External) is selected, use a signal
coming into the standard frequency input/output port on the back as the 10 MHz
standard frequency.
I/O DETECT >> 10 MHz
■ Detect
It enables you to select among Peak, Sample, Normal, Average, and Neg Peak detection
modes.
▪ Peak
When selected, the peak is shown on the upper part of the left of the display. The detection of
the peak is mainly used for detecting a signal on the noise level.
Find out the maximum level between the current display indicating point and the next indicating
point, and save the level on a trace memory where it meets the current indicating point.
▪ Sample
When selected, the Samp is shown up on the upper part of the left of the display.
The detection of a sample is used for indicating noise or a signal similar to noise.
The inspecting signal level on the current display point is saved on a trace memory.
▪ Normal
When selected, the Norm is shown up on the upper part of the left of the display.
The detection of normal is the detection mode combining the accurate signal measurement
of the peak, and the noise display function of the detection of a sample.
52
▪ Average
When selected, the Avg is shown up on the upper part of the left of the display. It saves the
average data between the detection of the Peak and the Neg Peak. It is used for reducing
VBW or irregular noise without using the trace average function. This function makes the
indication of the averaged value to be with swifter sweep speed.
▪ Neg Peak
When selected, the NPk is shown up on the upper part of the left of the display.
The detection of the Neg Peak is used for searching for a similar signal below the noise level.
Find out the minimal level between the current display indicating point and the next
indicating point, and save the level on a trace memory where it meets the current indicating
point.
I/O DETECT >> Detect
53
5-7. Marker
You can turn on or off by selecting a marker type and number. The marker is a diamond-shaped letter
showing a point. There can be four couples of markers at the maximum on the display at the
same time. Only one couple can be controlled at one time. The marker controlled is called
“activated” marker. If you press Marker, the Normal menu key gets activated.
■ Select Marker 1 2 3 4
Select one marker among 4 markers. Then, a marker already on, is activated when
selected. When the marker is designated to a different trace while it is already on, the
marker will be inactivated on trace concerned when selected.
MARKER >> Select Marker
■ Normal
When the marker is not yet indicated, a single frequency marker on the currently
selected marker will be activated on the center frequency. When the marker is already
indicated, it becomes activated on a selected location. The marker number is indicated on
the marker. The indications on the activated function block and the right of the display
show the frequency and amplitude of a marker. If you press Normal, the Delta function
goes out, and the activated marker moves to the location of the Delta marker.
MARKER >> Normal
■ Delta
The second marker gets activated on the location of the first marker. (When there is no
marker, two markers appear on the center of the display.) The amplitude and frequency
of the first marker were fixed. The marker number is indicated on the Delta marker, and
the same number is indicated on the standard marker as R. (For instance, 1R) The
indications on the activated function block and the right of the display show the
difference between the frequency and amplitude of two markers.
MARKER >> Delta
54
marker number is indicated on the marker as R, (for instance, 1R) and Delta marker is
indicated as a marker number.
MARKER >> Delta Pair
■ Off
Turn off the selected marker with Select Marker 1 2 3 4 keys. It removes a marker
indication on the display.
MARKER >> Off
■ Readout
It changes the activated marker value.
▪ Frequency
It sets the marker value as frequency. It is a basic value when it is not on zero span,
The following keys are used.
▪ Period
It sets the marker value as a period. It shows the reverse value of the frequency.
55
▪ Time
It sets the marker value as a time. It is a basic value when it is on zero span.
MARKER >> Readout
56
5-8. Marker →
■ Mkr → CF
It sets the marker frequency as the center frequency. On Delta marker mode, Mkr → CF
sets the center frequency as the marker delta value. Mkr → CF cannot be used when it is on
zero span.
MARKER → >> Mkr → CF
■ Mkr → CF Step
It changes the scale of the center frequency step so that it can match the activated marker
value. In order to see the scale of the step, please press FREQUENCY >> CF Step. As
for marker delta mode, the scale of the step is established as the difference between a
marker and the frequency. Mkr → CF Step cannot be used when it is on zero span.
MARKER → >> Mkr → CF Step
■ Mkr → Start
It sets the marker frequency as the start frequency. On delta marker mode, Mkr → Start
sets the start frequency as the marker delta value. Mkr → Start cannot be used when it
is on zero span.
MARKER → >> Mkr → Start
■ Mkr → Stop
It sets the marker frequency as the stop frequency. On delta marker mode, Mkr → Stop
sets the stop frequency as the marker delta value. Mkr → Stop cannot be used when it
is on zero span.
MARKER → >> Mkr → Stop
57
■ Mkr Delta → CF Step
It sets the difference between markers as the scale of the center frequency. If you want to
see the scale of the step, please press FREQUENCY >> CF Step. Mkr Delta → CF Step
cannot be used when the marker is off, or is on zero span.
MARKER → >> Mkr Delta → CF Step
58
5-9. Meas Control
It enables you to temporarily stop the power measurement function usable on a measure
key menu, or resume it. Also, the Meas Control enables you to select continuous sweep or
single sweep.
■ Restart
It repeats measurement from the beginning.
MEAS CONTROL >> Restart
■ Pause
It temporarily stops measurement. If you press Pause, it makes conversion between the
temporary pause of measurement and resume.
The key label makes conversion between Pause and Resume.
MEAS CONTROL >> Pause
■ Resume
It resumes the temporarily suspended measurement. The key label makes conversion
between Pause and Resume.
MEAS CONTROL >> Resume
59
5-10. Meas Setup
It indicates the measurement set-up menu. The indicated set-up menu shows differently in
accordance with selected measurement (ACP, Channel Power, Occupied BW, etc.) on the
measure menu.
▪ Ch Integ BW
It designates the scope used in calculating the power on the center channel.
MEAS SETUP >> Ch Integ BW
▪ Offset BW
It designates the scope used in calculating the power on the adjacent channel.
MEAS SETUP >> Offset >> Offset BW
▪ Offset Freq
It designates the difference between the center frequency of the center channel, and the
center frequency of the adjacent channel.
MEAS SETUP >> Offset >> Offset Freq
60
MEAS SETUP >> Avg Number
▪ Integ BW
It sets the scope of making calculation of channel power.
MEAS SETUP >> Integ BW
▪ OBW span
It sets the span of the analyzer on the occupied bandwidth measurement.
MEAS SETUP >> OBW Span
▪ Occ BW % Pwr
It changes the ratio of signal power used for determining the occupied bandwidth.
MEAS SETUP >> Occ BW % Pwr
61
5-11. MEASURE
■ Meas Off
It turns off the measurement function.
MEASURE >> Meas Off
■ ACP
It calculates power in the center and adjacent channel of a signal. A marker point can be
established as a measurement set-up menu while ACP is selected.
If you press Meas Setup while ACP is selected, you can set the adjacent channel power
measurement parameter. If you press Meas Control when ACP is selected, the
measurement can be temporarily suspended or resumed or makes conversion between
continuous or single modes.
MEASURE >> ACP
■ Channel Power
It calculates power and power spectrum density on a channel bandwidth designated by a
user. A marker on the display indicates the end of the channel bandwidth. The marker
selects Channel Power, and can be established as a Meas Setup menu. It operates on
single or continuous sweep mode.
If you press Meas Set-up after selecting Channel Power, you can set the channel power
measurement parameter. If you press Meas Control after selecting Channel Power, you
can temporarily suspend or resume measurement or make conversion between
continuous or single sweep modes.
MEASURE >> Channel Pwr
■ Occupied BW
The basic value of measurement is 99% of occupied bandwidth power. The measurement
operates on single or continuous sweep mode.
62
If you press Meas Set-up after selecting Occupied BW, you can set the occupied
bandwidth power measurement parameter. If you press Meas Control after selecting
Occupied BW, you can temporarily suspend or resume measurement or make conversion
between continuous or single sweep modes.
MEASURE >> Occupied BW
63
5-12. Preset
It sets the state of the analyzer to the state of being delivered from a factory, while
pressing Preset (when Preset (Factory) is selected). As for the set-up of the state of a free set,
please refer to the following table of conditions of free set. If you are to select Preset
(Factory), press SYSTEM >> Power On/Preset >> Preset Type (Factory).
Attenuation 30 dB
Start Frequency 0 Hz
Detect Normal
Frequency Offset 0 Hz
Marker Off
Sweep Continuous
Title Clear
Trace 1 Clear-Write
Trace 2 Blank
Trace 3 Blank
64
5-13. Print
Please press a print key if you want to immediately output the screen with the currently
defined printer. The screen will remain temporarily suspended until the data transmission is
completed.(there is no more sweep proceeded.) As for the set-up of a printer, please refer to
the explanation of a Print Set-up key in this chapter.
■ Printer
Determine whether to output printed data via a Centronics printer port or USB printer port.
PRINT SETUP >> Printer
■ Printer Setup
Select a printer driver concerned on the inside of the analyzer.
PRINT SETUP >> Printer Setup
■ Orientation
Select between Portrait or Landscape print.
PRINT SETUP >> Orientation
■ Page Size
Select a paper to print.
PRINT SETUP >> PageSize
65
5-15. Return
It returns to the previous menu. If you repeatedly press this key, it will return to the menu
previously selected.
It moves a marker to the maximum label by detecting the largest label among waves
outputted from the current screen, in the entire trace. If there is no peak, the marker will not
move. When there is no mark indicated, marker 1 will appear.
■ Next Peak
It detects the second largest peak next to the current marker label, and moves it to the
place. If you successively performs it, you can detect the peak that you can find the next,
and move the marker to each peak.
PEAK SEARCH >> Next Peak
■ Next PK Right
It moves to the next peak of the right of the current marker. In order to be recognized as
the signal peak, the level of the signal should be moved after it increase or decrease as
much as the value of excursion (6 dB). When there is no peak on the right, the marker
does not move.
PEAK SEARCH >> Next PK Right
■ Next PK Left
It moves to the next peak on the left of the current marker. In order to be recognized as the
signal peak, the level of the signal should be moved after it increase or decrease as much as
the value of excursion (6 dB). When there is no peak on the left, the marker does not move.
PEAK SEARCH >> Next PK Left
■ Min Search
It moves a marker on the current trace to the minimal level.
PEAK SEARCH >> Min Search
66
■ PK-PK Search
It finds out and indicates the difference between frequencies between the highest tracer
point and the lowest tracer point (or when the span is 0), and the amplitude.
PEAK SEARCH >> PK-PK Search
■ Continuous Pk On Off
When it is On, the marker is located on the highest level of trace, and when the sweep is
over, it successively detects the highest level, and locates the marker.
PEAK SEARCH >> Continuous Pk
■ Mkr → CF
It sets the center frequency of the analyzer as the marker frequency. On delta mode, Mkr →
CF sets the center frequency as the marker delta value. Mkr → CF cannot be used when it is
on zero span.
PEAK SEARCH >> Mkr → CF
67
5-17. Single Sweep
When the analyzer is on continuous sweep mode, it converts to a single sweep. When the
analyzer is already on single sweep mode, it performs a new sweep once if the trigger meets
the conditions.
■ SINGLE
SWEEP >> Sweep
5-18. SPAN
It activates the span function, and makes it possible to change the span.
■ Full Span
It sets the span as the entire frequency scope of the analyzer.
SPAN >> Full Span
■ Zero Span
It changes the frequency span as 0.
SPAN >> Zero Span
■ Last Span
It changes the frequency span of the analyzer to the previous span.
SPAN >> Last Span
68
5-19. Sweep
69
5-20. System
■ Power On Preset
If you press the Power On Preset button, the following menus will appear.
Although you press Preset, the set-up (Last or Preset) of the function of turning on power is
not changed, To change the set-up of the state of the analyzer restored when inputting power,
please use the Power On/Preset menu key.
■ Time/Date
There appears a menu that you can change time and date.
▪ Time/Date
It turns on and off the time and date.
70
SYSTEM >> Time/Date >> Time/Date
▪ Date Format
It changes the display type of a date from MM-DD-YY to DD-MM-YY.
SYSTEM >> Time/Date >> Date Format
▪ Set Time
It sets the time. Please input the time in accordance with the 24 hrs HHMMSS form by using
a number key pad, and press Enter. The effective time values (HH) are 00 ~ 23. The effective
MM and SS values are 00 ~ 59.
SYSTEM >> Time/Date >> Set Time
▪ Set Date
It sets a date. Please input a date in accordance with YYYYMMDD form by using a number
key pad, and press Enter. The effective YYYYs are from 2000 to 2037. The effective MM
value is 01 ~ 12, and the DD is 01 ~ 31. Use the following keys.
SYSTEM >> Time/Date >> Set Date
■ Alignments
It enables you to accurately acquire frequency and level by arranging the internal circuit
of the analyzer.
▪ Align Now
It immediately starts the arrangement of a system.
SYSTEM >> Align Now
▪ Align Auto
Off turns off the automatic arrangement. When you select On, the arrangement work of the
analyzer will be automatically made every 30 minutes.
SYSTEM >> Align Auto
■ Show System
There are the model number, serial number, firmware version, the version of each board
within the device, and the state of the set-up of an option of the analyzer.
71
SYSTEM >> Show System
■ Diagnostics
■ Program Update
LPT-3000 offers convenient ways to update firmware. Copy firmware update file
downloaded from a corporate website on a USB flash memory. Equip the USB flash
memory to the front USB port, and press a menu key concerned. Then you will see a file
list, and then press an update menu key while placing a file selection bar on an update
file. Then, there will be a message of “Program update success. Reboot LPT-3000.” On a
status window on the most lower part of display LCD, when the update was successful,
and if you turn on the power of the analyzer and turn on it again, the updated file will be
applied to the analyzer.
SYSTEM >> Program Update
72
5-21. Trig
■ Free Run
When sweep is over, it will immediately start the next sweep.
TRIG >> Free Run
■ Video
It is synchronized with an ascending angle of detection wave bigger than trigger level,
thus, it will start sweep.
TRIG >> Video
■ Line
It is synchronized with AC power frequency, it will start sweep. Line trigger is
conveniently used when observing waves related to power.
TRIG >> Line
73
5-22. Trace
It saves and establishes amplitude information drawn on the display LCD screen. Each
trace is composed of a series of data memories including the amplitude information. The
analyzer renews trace information whenever the sweep is over.
■ Trace 1 2 3
Select trace 1, trace 2, and trace 3. Please press a trace key until there appears an
underline below the trance number that you want.
TRACE >> Trace
■ Clear Write
It deletes data previously saved in trace memory selected, it saves amplitude data newly
acquired after sweep, and indicate amplitude data on display LCD screen. This function
is applied to trace 1 when inputting power.
TRACE >> Clear Write
■ Max Hold
Keep the current amplitude data of each point of trace memory selected, and renew the
current trace memory with new amplitude data when the bigger level than the amplitude
data of the current trace memory among amplitude data newly acquired from sweep, is
acquired.
TRACE >> Max Hold
■ Min Hold
Keep the current amplitude data of each point of trace memory selected, and renew the
current trace memory with new amplitude data when the smaller level than the amplitude
data of the current trace memory among amplitude data newly acquired from sweep, is
acquired.
TRACE >> Min Hold
■ View
Keep the amplitude data of trace memory selected, and indicate it on display LCD. Trace
memory does not be renewed even after sweep. When inactivating trace by pressing
74
Blank, you can read trace data saved by pressing View.
TRACE >> View
■ Blank
Save amplitude data on the selected trace, and delete it on the display. The trace memory
selected will not be renewed even after sweep. This function will be applied to trace 2
and 4 when inputting AC power.
TRACE >> Blank
75
Chapter 6. Function Test
6-1. Definition of Function Test
The function test means a test on diverse functions of the analyzer, and is to check whether the
analyzer normally operates. Before inspection or after repair, it is better check the operation the
analyzer. The objective of the function test lies in testing a device operating on the scope of
stipulated temperature suited to the standard of the device with the minimal test device.
The following is the explanation on the test, thus, it lists up each test standard, explanation
on the test objective or measurement value, lists of the specifications of devices necessary for
the test performance, test configuration, and test procedure by step. In terms of the test
objective, the operation of the device should be tested in the scope of stipulated temperature
for operation, suited to the standard of the device. You should perform test only after the
stipulated heating time has passed.
79
6-3. Displayed Average Noise Level
■ Test Standard
[Figure 6-1]
80
■ Test Procedures (10 MHz ~ 1.0 GHz)
1. Please press the Preset of the analyzer. Wait until the Preset procedure is over.
FREQUENCY >> Center Freq >> 1.0 GHz
SPAN >> 20 kHz
AMPLITUDE >> -70 dBm
BW/AVG >> 1 kHz
Video BW >> 10 Hz
4. Please record the average noise level read in the display line.
81
3. Please press the following key on the analyzer.
DISPLAY >> Display Line (On)
Please ignore the remaining response, and adjust the display line to appear in the center of noise.
4. Please record the average noise level read in the display line.
4. Please record the average noise level read in the display line.
82
6-4. Frequency Readout Accuracy
■ Test Standard
Example of the frequency readout accuracy : When frequency 1 GHz, Span 100 kHz,
RBW 1 kHz,
83
■ Test Procedures
Please connect the device as the figure 6-2. Connect the 10 MHz REF OUT of the analyzer
to the 10 MHz REF IN of a signal generator.
[Figure 6-2]
1. Please set the control device as follows, after pressing INSTRUMENT PRESET of a signal generator.
FREQUENCY >> 1.0 GHz
AMPLITUDE >> -10 dBm
2. Please wait until the Preset procedures are over, after pressing the Preset of the analyzer.
Please set the analyzer by pressing the following keys.
FREQUENCY >> 1.0 GHz
SPAN >> 100 kHz
BW/AVG >> Res BW >> 1 kHz
Video BW >> 1 kHz
3. Please measure the frequency value by pressing the PEAK SEARCH of the analyzer. When the
device is normally operating, the measured frequency should be between 0.999996000 GHz ~
1.000004000 GHz.
84
5. Please measure the frequency value by pressing the PEAK SEARCH of the analyzer. When the
device is normally operating, the measured frequency should be between 0.999649500 GHz ~
1.000350500 GHz.
■ Test Standard
85
■ Test Procedures
Adjust the zeroing of the power meter and power sensor.
Please connect the device as the figure 6-3
[Figure 6-3]
2. Please wait until the Preset procedures are over, after pressing the Preset of the analyzer.
Please set the analyzer by pressing the following keys.
FREQUENCY >> 80 MHz
CF Step >> 50 MHz
SPAN >> 20 kHz
AMPLITUDE >> -30 dBm
AMPLITUDE >> -30 dBm >> Attenuation >> 0 dB
Scale/Div >> 2 dB
BW/AVG >> 10 kHz
Video BW >> 3 kHz
PEAK SEARCH
3. Please adjust the frequency of the signal generator so that the marker frequency value is between -
34 dBm + / - 0.10 dB.
86
4. Please refer to the table 6-1, Frequency Response Report. Please input 80 MHz of the marker
reading value on an amplitude column of the analyzer as indicated.
5. Please input the power meter reading value on an amplitude column of the power meter.
6. Calculate the 80 MHz frequency response error with the following formula, and record the result
on a response error column.
Frequency response error = Amplitude of the analyzer - Amplitude of a power meter
7. Please perform the following steps by set-up of the center frequency in the table 6-1.
1) Adjust source to the following frequency in the center frequency column.
2) Perform the calibration of the power meter to the new test frequency.
3) Adjust the center frequency of the analyzer by pressing FREQUENCY, Center Freq, “n”, MHz.
(here, “n” is the following test frequency in the table 6-1.)
4) Press PEAK SEARCH.
5) Input the reading value of the power meter on the power meter amplitude column.
6) Input the reading value of the analyzer on the amplitude column.
7) Calculate the frequency response error with the following formula, and record the result on a
response error column.
Frequency response error = Amplitude of the analyzer – Amplitude of the power meter
The frequency response error should be smaller than the designated scale.
[Table 6-1]
87
6-6. Reference Level Accuracy
■ Test Standard
[Table 6-2]
88
■ Test Procedures
Please connect the device as figure 6-4.
[Figure 6-4]
2. Please wait until the preset procedure is over, after pressing the Preset of the analyzer.
Please press SYSTEM >> Alignments >> Align Now. Set the analyzer by pressing the following keys.
FREQUENCY >> 80MHz
SPAN >> 50 kHz
AMPLITUDE >> -30 dBm
BW/AVG >> Res BW >> 1 kHz
Video BW >> 30 Hz
3. Please press the MARKER SEARCH of the analyzer. Make the marker amplitude of the analyzer
to be between -30 dBm +/- 0.10 dBm by adjusting the amplitude of the signal generator. Please
record the power level of the signal generator on the amplitude standard of the signal generator of
table 6-2.
4. Please adjust the power level of the signal generator and the reference level of the analyzer in accordance
with table 6-2, since the standard of the 4th step result was established. (The output power of the signal
89
generator and the reference level of the analyzer will be changed on 10 dB step.)
5. Please record data measured by the analyzer in accordance with the changes of the power level of
the signal generator and the reference level of the analyzer, on the marker reading value of the
analyzer of table 6-2.
[Table 6-2]
90
6-7. Resolution Bandwidth Switching Error
■ Test Standard
■ Test Procedures
1. Please wait until the Preset procedures are over, after pressing the Preset of the analyzer.
Please set the analyzer by pressing the following keys.
91
SPAN >> 50 kHz
AMPLITUDE >> -30 dBm
Scale/Div >> 2 dB
BW/AVG >> 1 kHz
Video BW >> 1 kHz
3. After pressing AMPLITUDE, adjust reference level until a signal appears below the reference
level by using a scroll knob. Normally, the signal is located on the center of display.
MARKER SEARCH
MARKER >> Delta
4. Please set the resolution bandwidth and span of the analyzer in accordance with table 6-3.
5. After pressing PEAK SEARCH, please record the △ Mkr 1 amplitude value on table 6-3.
6. As for the set-up of other resolution bandwidth and span shown on table 6-3, please repeat the 3rd
and 4th steps.
The △ Mkr 1 amplitude value should be between the scope shown on the test standard.
[Table 6-3]
92
6-8. Display Scale linearity
■ Test Standard
■ Test Procedures
Please connect the device as figure 6-5.
[Figure 6-5]
93
1. Please set the signal generator as follows.
FREQUENCY >> 80 MHz
AMPLITUDE >> -30 dBm
2. After pressing the Preset of the analyzer, please wait until the Preset procedure is over.
Press SYSTEM >> Alignments >> Align Now.
FREQUENCY >> 80 MHz
SPAN >> 10 kHz
AMPLITUDE >> Attenuation >> 30 dB
BW/AVG >> Res BW >> 1 kHz
Video BW >> 10 Hz
PEAK SEARCH
3. Please adjust the amplitude of the signal generator until the marker amplitude of the analyzer
moves between 0 dBm +/- 0.10 dB.
Signal Generator Level Minimum (dB) Marker Level (dB) Maximum (dB)
Standard = ____ 0 ( standard ) 0 ( standard )
Standard -10 dB
Standard -20 dB
Standard -30 dB
Standard -40 dB
Standard -50 dB
Standard -60 dB
Standard -70 dB Total -1.5 dB Total 1.5 dB
94
6-9. Second Harmonic Distortion
■ Test Standard
≤-60 dBc, -40 dBm input
■ Test Procedures
Please connect the device as figure 6-6.
[Figure 6-6]
95
1. Please set the signal generator as follows.
FREQUENCY >> 50 MHz
AMPLITUDE >> -40 dBm
2. Press the Preset of the analyzer, Wait until the Preset procedure is over.
Set the analyzer by pressing the following keys.
FREQUENCY >> 50 MHz
SPAN >> 1 MHz
AMPLITUDE >> -40 dBm >> Attenuation >> 0 dB
BW/AVG >> 30 kHz
3. Adjust the amplitude of the signal generator so that the signal peak gets to reference level.
5. Wait until two times of sweep are over, and press the keys on the analyzer as follows.
MARKER SEARCH
MARKER >> Delta
6. Please move the center frequency of the analyzer to the second harmonics (100 MHz).
Press Peak Search. The marker delta amplitude value should be below the test standard.
96
Chapter 7. Measurement Guide
7-1. Composition
The following examples of use explain the functions of the analyzer by taking a simple
example of measurement generally used. When an inexperienced user performs experiment,
he/she can promptly learn how to use.
99
7-2. Utilization of Delta Marker ; Comparison of Signals
You can easily compare the difference of frequency and amplitude between spectrum
signals of radio, wireless device, mobile communication phone, and CATV by using the
analyzer. If you use the delta marker function of the analyzer, you can compare the
difference of power between two signals.
1. Connect 10 MHz REF OUT on the back panel of the analyzer to RF INPUT on the front panel of
the analyzer.
2. By pressing FREQUENCY >> 30 MHz >> SPAN >> 50 MHz, set the center frequency to 30
MHz, and set the span to 50 MHz.
100
[Figure 7-1] Reading 10 MHz signal by using marker
6. Turn on the delta marker by pressing MARKER >> Delta, Move the delta marker to the next
peak by using a scroll knob or PEAK SEARCH >> Next Peak.
7. The differences of amplitude and frequency between markers appear on the right of the upper
part of display LCD.
101
7-3. Utilization of RBW ; Distinction of a Small Signal
In order to distinguish two signals with similar signal, you need to consider the feature of the form of the
analyzer IF(RBW) filter and 3 dB bandwidth. The feature of the form of a filter is defined by selectivity
which is a ratio of 3 dB bandwidth and 60 dB bandwidth. If a small signal is too near a big signal, the small
signal might be hidden by the IF(RBW) filter selection.
1. In order to gain a signal with 200kHz of frequency difference, connect the analyzer as figure 7-3.
Set the first signal generator to be 100 MHz, -30 dBm.
2. By pressing FREQUENCY >> 100 MHz >> SPAN >> 1 MHz, set the center frequency of the
analyzer to 100 MHz and set the span to 1MHz.
3. By pressing I/O DETECT >> Detect >> Peak, set detect mode as peak.
5. Set the second signal generator to 100.2 MHz, set the signal 200 kHz higher than the first signal.
Set the signal amplitude to -90 dBm. (it is lower than the first signal by 60 dB.)
102
[Figure 7-4] Signal Resolution with 10 kHz of RBW
6. By pressing PEAK SEARCH >> MARKER → >> Mkr → Ref Lvl, set the 100 MHz signal as
reference level. When a 10 kHz filter with less than 15 : 1 of selectivity is used, the bandwidth of
the filter will be less than 150 kHz on the point of 60 dB, and the half of the disintegration
capacity bandwidth becomes smaller than the difference of frequency (less than 75 kHz), you
can distinguish an input signal.
7. By pressing MARKER >> Delta >> PEAK SEARCH >> Next Pk Right, move the marker on
a small signal. When a filter with RBW 30 kHz is used, the area of 60 dB bandwidth will be less
than 450 kHz.
Since the half of bandwidth (less than 225 kHz) is larger than the difference of frequency (200
kHz), you cannot nearly distinguish a signal.
103
[Figure 7-5] Signal Resolution with RBW 30 kHz
8. Figure 7-6 is an example of using an RBW 100 kHz filter. It shows that a signal with 200 kHz
distance, enters the inside of RBW 100 kHz bandwidth, thus, any signal cannot be distinguished
at all. As such, you can distinguish a delicate signal by adjusting RBW.
104
7-4. Measurement of a Small Signal
The capacity of the analyzer for measuring a small signal is limited by noise level
occurred inside of the analyzer itself. A small signal with poor power level may not be seen
while hidden by noise level of the analyzer. As for sensitivity on the measurement of a small
signal, the set-up of the device is the most important when measuring, thus, noise level
within the analyzer get affected by the set-up state.
RBW set-up affects noise level within the analyzer the most, and the attenuator of input
part of the analyzer affects the level of a signal measured. Ultimately, you can accurately
measure a small signal if you set the analyzer with high S/N ratio (Signal - to - noise ratio).
(RBW, Atten Î set it relatively smaller) The following example shows several examples for
measuring a small signal. When it is difficult to distinguish a small signal from noise even
after setting RBW, attenuator, you can enhance the visuality by reducing video bandwidth
and using the video average function. The reduction of video bandwidth and average
function averages irregularly occurring noise.
2. Set the frequency of the signal generator to 100 MHz, and set the amplitude to -90 dBm.
105
7. Press AMPLITUDE >> Attenuation (Man). By pressing a step key, select 10 dB attenuation
rate, If you increase the attenuation rate, noise will move closer to a signal.
106
■ Examples of Use (RBW Selection)
By reducing RBW and lowering noise level of the analyzer, you can measure a small signal.
1. Set the frequency of the signal generator to 100 MHz, and set the amplitude to -90 dBm.
6. After pressing BW/AVG >> RBW (Man), decrease the disintegration capacity bandwidth by
pressing a step key. Since noise level of the analyzer decreases, a small signal will appear more
clearly.
107
[Figure 7-10] Signal with RBW 3 kHz
There is # mark in front of RBW mark on the lower left of the screen, which means it is not
RBW Auto. As RBW decreases, the sweep time will increase.
Measure the level of a small signal by using the video bandwidth function.
1. Set the frequency of the signal generator to 100 MHz, and set the amplitude to -90 dBm.
108
5. Perform ten times of trace average.
6. After pressing BW/AVG >> VBW (Man), decrease the disintegration capacity bandwidth by
pressing a step key. Since noise level of the analyzer decreases, a small signal will appear more
clearly and the measurement of signal level will improve.
109
[Figure 7-13] Signal with VBW 1 kHz
110
[Figure 7-15] Signal with VBW 100 Hz
The final result of trace average is similar to the reduction of video bandwidth. As video
bandwidth reduces, the sweep time increases, and the trace average will use the
111
averaging of many sweep data.
1. Set the frequency of the signal generator to 100 MHz, and set the amplitude to -90 dBm.
5. By pressing BW/AVG >> Average (On), operate the trace average. If the trace average function
softens trace, the visibility of a small signal will be enhanced.
6. In order to set the number of sweep, input the number by using a number key while the average
button is pressed. After inputting the number, if you press the next unit key, the averaging of
trace will be initialized, and resumed.
While the trace average function is performed, the current sweep counter will appear on the
center of the left of the screen. When the sweep with established time is completed, the analyzer
will keep trace averaging based on the established number.
112
7-5. Measurement of CDMA Channel Power
Channel power measures the power of the area equivalent to the channel bandwidth
designated on the base of the current center frequency.
1. Set the frequency of CDMA signal generator to 850 MHz, and set the amplitude to -30 dBm.
2. By pressing MEASURE >> Channel Pwr, change the analyzer to the power measurement mode.
5. Set Integration BW to 2 MHz. Then, the entire span will be automatically established as 1.5
times of Integration BW.
6. By pressing MEAS SETUP >> Avg Number >> 10 >> ENTER, perform ten times of average.
113
7-6. CDMA ACP Measurement
1. Set the frequency of CDMA signal generator to 850 MHz, and set the amplitude to -30 dBm.
2. By pressing MEASURE >> ACP, change the analyzer to ACP measurement mode.
8. By pressing MEAS SETUP >> Avg Number >> 10 >> ENTER, perform ten times of average.
114
7-7. Printer Use
On LPT-3000, it offers USB port apart from a parallel port for the connection to a printer.
Set a port to transmit print data on a printer of PRINT SETUP menu. As for USB port on the
front, you can save data by using USB memory apart from a print port.
■ Examples of Use
1. On PRINT SETUP >> Printer, select a parallel port or USB port.
4. On PRINT SETUP >> Page Size, select the size of printing paper.
115
7-8. USB Memory Use
On LPT-3000, in order to bring picture data saved outside of the analyzer, USB flash
memory is used instead of 3.5″ floppy disk. As for locations to save data of LPT-3000, there
are “screen” which is an internal memory of the analyzer, and “usbfs” indicating USB flash
memory. If you set a path to usbfs on the file list screen, while USB flash memory is
attached to USB port on the front of the analyzer, operations including Save, Load, Delete,
Rename, and Create Dir will perform not as a screen, which is an internal memory of the
analyzer, but as USB flash memory on File menu. Also, if you get out of the file menu after
setting a path to usbfs on the file list screen of the file menu, the saving of a file by the
operation of Save key on the front board will be directly performed not as an internal
memory of the analyzer, but as USB flash memory.
■ Examples of Use
1. Equip USB flash memory to USB port on the front.
2. By pressing FILE >> Save >> Type >> Screen keys, set a file type to a picture file.
3. By pressing FILE >> Save >> Format >> GIF keys, set a picture format file.
4. By pressing FILE >> Save >> Dir Select keys, set a path to usbfs.
5. By pressing RETURN key, comes out to the analyzer wave sweep screen.
6. Press SAVE key. Then, the current display screen will be saved as a picture file on USB flash
memory.
116
7-9. Trace Use
It offers three trace memories on LPT-3000. The movement between trance memories will
be used by the first menu key.
Clear Write deletes the current data in trace memory, and rewrite data newly acquired
after sweep, on the memory, This operation is successively performed after every sweep is
over. Clear Write is the basic trace mode of the analyzer.
Max Hold saves the biggest value by comparing data in the current trace memory and
data newly acquired after sweep, and display them.
Min Hold saves the smallest value in trace memory by comparing data in the current
trace memory and data newly acquired after sweep, and display them.
View indicates the contents of the current trace memory while suspending the writing on
the display screen.
Blank deletes trace on the display screen. However, the contents of the trace memory do
not get deleted, and they will show up on the display screen, when View menu key is
pressed.
117
Chapter 8. Options